BIOGAS 5000 gas analyser
Page
58
Copyright Geotech
9.7.1
Gas Check in fresh air
START
Switch on the
instrument and
complete the
instrument warm up.
Main Gas Read
Screen select the
‘Menu’ key
Select key 2
‘Gas Check’
Select key 1
‘Gas check ’
Wait for the readings to
stabilise. Wait for the
analyser to zero for
approx. 1 min.
Message will display
‘Please wait for timer
to reach zero’
When complete a
message will
display ‘Check
complete’
Read status field
on the right hand
side of the screen
Select soft-key
‘Calibrate’
Read status field
on the right hand
side of the screen
Calibrate
complete?
Calibration check
ok?
Result saved
END
NO
Select
soft-key
‘Exit’
YES
NO
Select ‘Fresh Air’ by using
the enter key
Select soft-key
‘Retry’
Contact Geotech
technical support team
for advice
YES
Operation Settings
screen
Select
soft-key
‘Check’
Disconnect all
tubing
Exit
Calibrate complete
YES
Let the pump run for 90
seconds and select soft key
‘Check’