27
8. Apply the recommended simulation signal to
simulate 10 pH for your input card type
(Table 1). Repeat step 6 to enter and save the
second simulation value into memory.
9. Access main pc board and
open
dip switch
position 2 to disable the introduction mode
(sec. 4.1). This completes the introduction
mode procedure.
Optional: To enable the security feature,
close
dip switch 1.
You must now recalibrate your pH system. See
sections 3.2 - 3.3 for details.
0 0 0 0
p H > 1 0
0 0 . 0 0
p H > 1 0
0 0 0 0
p H > 1 0
OR
OR
pH Sensor (mV)
Input Card Display
4 to 20/0 to 20 mA
Input Card Display
0 to 5/0 to 10 VDC
Input Card Display
mV's displayed
mA's displayed
mV's displayed
7. Press:
CAL
(once). The 9030 reads:
Содержание SIGNET 9030 Intelek-Pro