Geometrics SmartSeis Скачать руководство пользователя страница 4

SmartSeis

TM

 Maintenance Manual 

 

4

 

 

During initial set-up, check the cable connectors for debris (dirt, dust) on, or damage to, the 
connecting surfaces, and firmly connect them to the seismograph.  If the seismograph displays a 

dead or noisy trace(s)

, there is likely a problem with the associated geophone or spread 

cable.  First, swap the spread cable connectors that plug into the seismograph and see if the 
problem moves to a different channel.  If the problem moves with the cable, the cable and/or 
geophone are suspect.   

Next check the geophone condition, its connection with the spread cable, and whether it is 
planted in the ground.  Also, try substituting a different geophone.  After these tests, if the cable 
is suspect, try reversing it or swapping it out.    

If the problem persists on a single channel and does not move with the cables or geophones, it 
could indicate that there is problem with the associated acquisition board.  See 

Section 2.2, 

Replacing an acquisition board

An entire 

set of 12 dead or noisy traces

 may also be attributed to cable condition or a bad 

board, as well as close cable proximity to a noise source.  Check the externally connected 
cables as described above.  If the cables are not suspect, remove the seismograph chassis from 
the yellow housing (see 

Section 2.1, Internal seismograph access

) and check the cable 

connections, starting with the panel connector(s), leading to the acquisition board(s).   

You can try swapping the connections to the acquisition boards if you have a 24-channel 
system.  If the problem is still on the same set of channels, then it is likely that the acquisition 
board is bad.  Replace it if a spare is available, referring to 

Section 2.2, Replacing an 

acquisition board

If noise is the problem, shift the cabling to see if that changes the nature of the noise.  If there is 
a change, this would indicate an external noise source too near the cabling. 

 

1.3 Triggering 

The seismograph can be triggered by contact closure, contact opening, or signal input, such as 
from a geophone.  The input signal level should never exceed 5 volts.  

If the system 

fails to trigger

, test whether the hammer switch is broken by triggering it 

connected directly to the seismograph, without the extension cable.  To trigger, tap the end of 
the brass cylinder on a table or hard object.  The seismograph should trigger each time and not 
false trigger prior to tapping of the cylinder.   If there is no trigger, swap out the suspected 
switch with a new one and retest it to confirm that it functions.  If you are using another device 
to trigger the seismograph, test it similarly for proper operation.   

If the system still fails to trigger, check if the seismograph trigger circuitry is operational.  Set the 
trigger sensitivity to the midpoint at 50.  Next, trigger the seismograph using the MANUAL 
TRIGGER function in the OTHER menu.   

Содержание SmartSeis

Страница 1: ...SmartSeisTM Maintenance Manual Geometrics Inc 2190 Fortune Drive San Jose CA 95131 Tel 408 954 0522 Fax 408 954 0902 Email sales mail geometrics com Internet www geometrics com...

Страница 2: ...1 Software 5 1 4 2 Hardware 6 2 0 SmartSeisTM Maintenance and Repair 9 2 1 Internal seismograph access 9 2 2 Replacing an acquisition board 10 2 2 1 Acquistion board jumper settings 11 2 3 Replacing t...

Страница 3: ...t the DOS prompt then press ENTER 1 1 External power The most common problem is power failure normally a discharged battery The front panel has a series of lights which constantly display the battery...

Страница 4: ...eck the cable connections starting with the panel connector s leading to the acquisition board s You can try swapping the connections to the acquisition boards if you have a 24 channel system If the p...

Страница 5: ...ed If false triggering persists it may be necessary to add external attenuation or isolation to correct the problem 1 4 Internal system 1 4 1 Software Since the SmartSeisTM is a normal PC compatible s...

Страница 6: ...ing the sun and is corrected when moved away and allowed to cool Storage devices Most problems with the floppy and hard drives will be indicated by error messages upon the system s self check during b...

Страница 7: ...plying the system components with a consistent 12 V Each LED indicates that a specific power module on the power supply board is working Also check the LEDs of the system components to confirm that th...

Страница 8: ...ent with the system 6 Reset the parameters that are different than the original settings Exit with saving If the changes that cause the errors occur again this is an indication that there is a problem...

Страница 9: ...ard s The boards are of a dense surface mount construction and they themselves do not contain parts that can be serviced in the field Repair of the boards not only requires specialized training and fa...

Страница 10: ...e power distribution cable from connector J2 the 10 pin ribbon cable located on the left and topsides of the board Figure 2 4 Disconnect the geophone ribbon cable the 26 pin ribbon cable located on th...

Страница 11: ...of 12 channels Depending on whether it is a 12 or 24 channel SmartSeisTM S12 or S24 the jumper installation will vary Table 1 outlines the jumper settings specific to either case Spare boards are ship...

Страница 12: ...revent wrong connections we suggest that you make a careful drawing of the cable connections specific to your seismograph prior to board removal There is one pair of connectors that are the same size...

Страница 13: ...5 Floppy drive J6 IDE hard drive J7 Power distribution board J8 Speaker port on motherboard J9 Not used J10 Trigger To remove an OMNI board 1 Confirm that the power is off and the cable disconnected f...

Страница 14: ...onnector pins or nearby parts Also each replaced board needs to be fully seated into its backplane slot reslotted into the bracket and squarely aligned with the chassis and the other boards As mention...

Страница 15: ...appropriate settings Table 6 outlines the setting for switch S1 Table 3 Jumper numbers and corresponding IRQ numbers used by the seismograph trigger Jumper JP4 and IRQ 11 is the normal selection Jump...

Страница 16: ...d because the cost of repair exceeds its purchase price problems are fixed by substitution If you do not have a spares kit you may be able to purchase a replacement floppy drive locally Press the silv...

Страница 17: ...the power cable is disconnected from the seismograph Refer to Section 2 4 Replacing the floppy drive in order to slide the floppy drive from its keyed slots that the hard drive bracket will be cleared...

Страница 18: ...se the LCD panel and adapter board Figure 6 Lift out the LCD panel and adapter board and disconnect the associated cables Installation of a new LCD panel is the reverse of removal Figure 6 The front L...

Отзывы: