![Geokon 6300 Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/geokon/6300/6300_instruction-manual_2214576016.webp)
10
3.3 GK-403 Readout Box (Obsolete Model)
The GK-403 can store gauge readings and apply calibration factors to convert readings to
engineering units. The following instructions explain taking gauge measurements using Mode
“B”. Consult the GK-403 Instruction Manual for additional information.
3.3.1 Connecting Sensors
Connecting sensors with 10-pin connectors:
Align the grooves on the sensor connector (male), with the appropriate connector on the
readout (female connector labeled senor or load cell). Push the connector into place, and
then twist the outer ring of the male connector until it locks into place.
Connecting Sensors with Bare Leads:
Attach the GK-403-2 flying leads to the bare leads of a Geokon vibrating wire sensor by
connecting each of the clips on the leads to the matching colors of the sensor conductors,
with blue representing the shield (bare).
3.3.2 Operating the GK-403
1)
Turn the display selector to position “B”.
2)
Turn the unit on.
3)
The readout will display the vibrating wire output in digits. The last digit may change
one or two digits while reading.
4)
The thermistor reading will be displayed above the gauge reading in degrees
centigrade.
5)
Press the “Store” button to record the value displayed.
If the no reading displays or the reading is unstable, see Section 5 for troubleshooting
suggestions. The unit will turn off automatically after approximately two minutes to
conserve power.
3.4 Measuring Temperatures
Each Vibrating tilt sensor is equipped with a thermistor for reading temperature. The thermistor
gives a varying resistance output as the temperature changes. Geokon readout boxes will read the
thermistor and display temperature in
°
C automatically. To read the thermistor using an
ohmmeter, complete the following:
1) Connect the ohmmeter to the two thermistor leads coming from the instrument. (Usually
white and green.) Since the resistance changes with temperature are large, the effect of cable
resistance is usually insignificant.
2)
Look up the temperature for the measured resistance in Table 2 in Appendix B.
Содержание 6300
Страница 2: ......
Страница 4: ......
Страница 24: ...18 APPENDIX E TYPICAL CALIBRATION REPORT Figure 11 Sample Model 6300 Calibration Report ...