-17
-
Temperatures should be recorded at the time of each reading
along with notes
concerning the construction activity that is taking place. These data might supply logical
reasons for observed changes in the readings.
8. TROUBLESHOOTING
Maintenance and troubleshooting of Model 4000 Vibrating Wire Strain Gages are confined
to periodic checks of cable connections and maintenance of terminals. Once installed, the
gages are usually inaccessible and remedial action is limited.
Consult the following list of problems and possible solutions should difficulties arise.
Consult the factory for additional troubleshooting help.
Symptom: Strain Gage Readings are Unstable
Is the readout box position set correctly? If using a datalogger to record readings
automatically are the swept frequency excitation settings correct?
Is the strain readings outside the specified range (either compressive or tensile) of the
instrument? Gage may have become too slack or too tight; inspection of the data might
indicate that this is a possibility. Loosen the two oval point set screws in one of the
mounting blocks. This will permit the internal spring to re-tension the gage and the gage
will read again. Set the gage to some new datum and retighten the setscrews. If the
gage does not respond to re-setting and if the old plucking coil will pluck a new gage,
then the gage should be replaced.
Is there a source of electrical noise nearby? Most probable sources of electrical noise
are motors, generators and antennas. Move the equipment away from the installation or
install electronic filtering. Make sure the shield drain wire is connected to ground
whether using a portable readout or datalogger.
Does the readout work with another gage? If not, the readout may have a low battery or
be malfunctioning.
Symptom: Strain Gage Fails to Read
Is the cable cut or crushed? This can be checked with an ohmmeter. Nominal
resistance between the two gage leads (usually red and black leads) is 180
,
10
.
Remember to add cable resistance when checking (22 AWG stranded copper leads are
approximately 14.7
/1000' or 48.5
/km, multiply by 2 for both directions). If the
resistance reads infinite, or very high (megohms), a cut wire must be suspected. If the
resistance reads very low (
100
) a short in the cable is likely. Splicing kits and
instructions are available from the factory to repair broken or shorted cables. Consult
the factory for additional information.
Does the readout or datalogger work with another strain gage? If not, the readout or
datalogger may be malfunctioning.