Service Manual FL 40-CrossPointer SP 

and the horizontal reference line are parallel, i.e. parallel error, please calibrate as the follow methods: 

Note: observe whether there is error between the horizontal laser line and the adjacent light  ruler reference. 

If there is error, please calculate the error into the reference on the wall. (if the adjacent light ruler laser line 

locates on +2, the laser line on wall reference locates on +2 too, it means the laser line is in horizontal in this 

direction, there is no need calibration.) 

 

2.2.1.1.1 If the excursion is not more than 10mm, adjust the adjusting screw in two self check hole on point A 

on the housing of the instrument (figure 3), make the horizontal laser line is coincident with the reference 

line. 

 

 

 

 

 

 

 

 

 

                                                                                    Figure 3 

2.2.1.1.2 If the excursion is more than  3mm, disassemble the housing parts

disassemble the screws at B

figure  3

,add  weight  stack

when  horizontal  line  is  lower  than  the  reference

,remove  the  weight  stack

when the horizontal line is higher than the reference

keep the horizontal line in a distance range of 10mm 

to the reference line, and adjust 3 fine adjusting screw at C finally to make the horizontal line superposition 

with the reference line.   

 

2.2.2 Gradient error 

As showed in Fifure 4, the laser line is incline to the reference line, this kind of error called gradient error       

 

 

 

 

 

Содержание Selection PRO FLG 40-PowerCross GREEN SP

Страница 1: ...FL 40 PowerCross SP Service Manual...

Страница 2: ...instrument 2 Accuracy calibration 2 1 Setting the reference 2 2 Sort of the error and calibration methods 3 Adjustment for alarming range 4 Disassembly and assembly of the instrument 4 1 Disassembly...

Страница 3: ...The hairspring is broken 3 The 1 controling PCB is broken 1 Jointing the broken connecting line 2 Change the broken hairspring 3 Change 1 circuit board Laser line is dim The laser diode is broken Cha...

Страница 4: ...adjacent light ruler to the right front of the adjusting platform the 0 position Of the adjacent light ruler should be consistent with the horizontal reference 0 position 2 2 Sort of the error and cal...

Страница 5: ...10mm adjust the adjusting screw in two self check hole on point A on the housing of the instrument figure 3 make the horizontal laser line is coincident with the reference line Figure 3 2 2 1 1 2 If...

Страница 6: ...figure 3 When adjust the two screws on point D it need to loosen one screw before tighten another screw until the horizontal laser line is coincident with the reference line The degrees of tightness o...

Страница 7: ...one screw before tighten another screw until the horizontal laser line icoincident with the reference line The degrees of tightness of the two screws should be nearly the same finally 2 2 3 2 Calibrat...

Страница 8: ...3 13 Then move 3 connecting PC board in horizontal make the alarming pole of 2 alarming PC board 3 13 locate in the centre of the alarming ring then fasten the screws When the incline degree of any di...

Страница 9: ...th a cutter 1 1 2 Use a cross screwdriver to turn off 4 pcs cross head plate tapping screw ST2 2x5 1 4 and take off 1 operation PC board G609 1 operation PC board 1 3 3 Pull a little harder to move of...

Страница 10: ...1 by a cross srewdriver move off 4 alarming PCB G609 4 alarming PCB 3 4 6 Screw off 1 pc of cross head plate screw M2X10 3 9 by a cross srewdriver move off horizontal laser module green laser module...

Страница 11: ...ery cover part 4 9 3 Use electric iron to burn down the leading line of the power supply plug move off the contacting part 4 6 4 Screw off 3 pcs of nut M2 5 4 2 by a cross srewdriver move off 6 pcs of...

Страница 12: ...board 6 15 6 1 Lifting bracket 6 2 Locking sheet 2 6 3 Damping bracket 6 4 Cross head plate tapping screw ST2 2X5 5 6 5 Locking ring 5 6 6 Cam 6 7 Cross head plate tapping screw ST2 9X9 5 9 6 8 Magne...

Страница 13: ...the Locking ring 6 6 17 saperate turnplate 6 22 2 pcs of gasket 6 18 2 pcs of saddle pad 6 19 lifting bracket 6 1 7 Screw off 4 pcs of cross head plate tapping screw ST2 9X9 5 6 7 move off the base co...

Страница 14: ...showed in figure 11 connect the corresponding connecting line to each PCBs accordingly Solder 4 hairsprings to the corresponding place of4 Alarming PC board or G609 4 Alarming PC board 3 4 and 3 Conne...

Страница 15: ...that note especially Requirement of assembly a Gelatinize 706 glue water on the jointing as follows The interface between base 4 5 and inserting block 4 13 Lifting bracket 6 1 and press cover 6 21 Fin...

Отзывы: