Error codes 

The following table describes the possible error codes displayed on the Error List of GEO 1000 SP 

 

 

 

Error String 

Control behavior 

 

“No Analog Detected” 

Control is turned off 

Internal Hardware fault 

“No Peltier” 

Control is turned off  

Internal over temp probe 

fault fans will run 

continuously. 

“NO HumGen Temp” 

Control is turned off 

Internal temperature probe 

failure. 

“Peliter High Limit” 

Control is turned off if halt on 

errors is enabled 

Internal cooling sensor is 

above limit 

“HumGen Low Limit” 

Control is turned off if halt on 

errors is enabled 

Ambient temperature of 

humidity generator is below 

operating temperature 

“HumGen High Limit” 

Control is turned off if halt on 

errors is enabled 

Humidity generator is above 

allowed limit. Check water 

supply or ambient conditions 

 

“EEPROM Cfg Reset” 

Control not affected 

Unit has reset its 

configuration to factory 

default review settings. 

“EEPROM Cal Reset” 

Control not affected 

Factory calibration has been 

reset to default. Please 

contact GEO instruments. 

 

 

 

Содержание 1000 SP

Страница 1: ...Copyright 2015 All Rights Reserved http www geocalibration com 1 631 471 6157 GEO Calibration 2190 Smithtown Ave Ronkonkoma NY 11779 USA GEO CALIBRATION Model 1000 SP User operational guide 10 15 2015...

Страница 2: ...tion prevention 8 Front Panel Operation 9 Main screen 9 Button Operation 10 Exit warming mode 10 Status Lights 11 Setting the current temperature and humidity 11 Internal Probe Calibration Configurati...

Страница 3: ...ii GEO Windows Application Usage 16 Main Screen 17 Experiments 17 Phase 18 Modbus Initial setup 19 Panel Mode 20 Plot control 21 Reports 21 Record Mode 21 Error codes 22 Contact Information 23...

Страница 4: ...Quick Start Guide Filling the Water Reservoir Locate the Fill Port Locate the Fill Tube Remove Cover from Fill Port A fill tube is shipping standard with every purchase of a GEO Calibration calibrato...

Страница 5: ...se standard distilled water Fill one 10 mL syringe at a time while watching the fill level The water will naturally flow into the internal chamber reservoir Due to the reservoir s 50 mL capacity do no...

Страница 6: ...or other solid object may be placed under the front of the unit for stability Locate Remove Drain Cap The water will immediately begin flowing from the unit into the bucket below WARNING Failure to re...

Страница 7: ...4 Cleaning Instructions Open the Chamber Door Unscrew any visible fasteners and pull the door from the chamber with moderate force Clean With Paper Towel Disinfect with Alcohol Wipes...

Страница 8: ...Locate the Two Bungs on Rear Make sure there are O Rings inside of each bung Push Desiccant into Bungs Make sure blue desiccant is on the left hand side Using gentle force press the desiccant into th...

Страница 9: ...val Unjoin Velcro Straps Loosen Velcro Straps Gently Pull Desiccant Pull with two hands using even force on both sides of the desiccant canister Use thumbs to push out from the chamber body as shown i...

Страница 10: ...push a white filter to the bottom Empty the contents of the white desiccant bottle into the canister Gently tap to pack Insert a second white filter Empty the contents of the blue desiccant bottle in...

Страница 11: ...000 SP allows the user to generate a stable environment within its chamber The unit can control and maintain the relative humidity and temperature in the chamber to within a resolution of 0 1 RH and 0...

Страница 12: ...ature and humidity inside the chamber From this screen the user can change the set point The screen is organized as a grid The grid has rows T and H which stand for Temperature and Relative Humidity a...

Страница 13: ...ced in edit mode Pressing the UP button increases the value of the field while pressing the DOWN button decreases the value Holding either the UP or DOWN button for more than one second will advance t...

Страница 14: ...he BIT Built in Test has detected a condition which will not allow control of the chamber See appendix Error code for the possible fault conditions Setting the current temperature and humidity From th...

Страница 15: ...1000 SP Heat exchanger removes the heat generated when the peltier element is cooling the chamber To prevent damage the unit monitors the temperature of the heat exchanger and will stop control if the...

Страница 16: ...recommended to change these values as they are pre configured for the optimal operation of the unit The thresholds are nevertheless configurable as shown in the figure below Figure 7 BIT Humidity gene...

Страница 17: ...control probe channel Configure DFB probe types using GEO supplied configuration or user defined custom probe types Create multiple experiments each of which may have multiple phases where either tem...

Страница 18: ...ystem If Windows is unable to find the driver please follow the instructions provided with converter or go directly to the converter manufacturer website or contact GEO technical support at the addres...

Страница 19: ...16 GEO Windows Application Usage...

Страница 20: ...ow the experiment to be started stopped The user may also manually proceed to the next phase of the running experiment or rewind to the beginning of the previous phase of the current experiment In the...

Страница 21: ...s defined by a starting point end point increment and an update rate Once the phase is started the system will set the environmental conditions to the values in start it will increment the value at a...

Страница 22: ...communication port to which the unit is connected must be found If the system is powered on pressing the Auto Configure button will cause the GEO application to find the communication port that the un...

Страница 23: ...from the main screen button labeled Panel Mode Panel mode is only available if an experiment is not currently running In Panel mode the user may manually control the environmental conditions with the...

Страница 24: ...ort to be printed on the currently collected information To print a report press the Print Report button from the main screen The user will be prompted for their name title employee ID and any additio...

Страница 25: ...is turned off if halt on errors is enabled Internal cooling sensor is above limit HumGen Low Limit Control is turned off if halt on errors is enabled Ambient temperature of humidity generator is belo...

Страница 26: ...23 Contact Information GEO Calibration 2190 Smithtown Ave Ronkonkoma NY 11779 1 631 471 6157 http www geocalibration com info geocalibration com...

Отзывы: