GNF - 722 - 12 (0)                                                         Genicom Co., Ltd.                                                            A4(297

×

210

Fig. 6  Connect diagram of MG-06 

Red light 

Green light 

Red light 

ballast 

Back side of  MG-06 , Numeric is terminal Number. 

         5            4            3            2            1 

Ballast and light connection 

AC power 

Fig. 7  Wiring diagram of Green light and Red light 

(Red light means abnormal , Green light means normal) 

Power cable 

Power cable 

Green light 

ballast 

Capacitor 

220

/2kV 

Page 6 

In normal status , the relay will connect #6 and #7 , the Green light will be ON  and Red light will be off 
In abnormal status , the relay will connect #7 and #8 , the Green light will be of  and Red light will be ON. 

Sensor Probe 

AC Power 

Supply 

10 

         10            9            8            7            6 

6. Relative Reaction Curves of UV Sensors 

Fig. 4  Relative Responsivity Curve of UV Sensor 

(1) UVV Sensor 

(2) UVC Sensor 

Содержание MG-06

Страница 1: ...and Applications 2 Product Specification and Configuration 3 Product Installation 4 Operation 5 Relay Operation method 6 Relative Reaction Curves of UV Sensors 7 A S Request in Case of Product Failur...

Страница 2: ...Configuration 1 Display Panel Panel Size 48 48 85 5 Panel cutting size 45 45 Tolerance 0 0 5 Power Supply 85 265 VAC 50 60 Consumption 3 W Operating Temperature 0 50 Operating Humidity 35 85 RH Relay...

Страница 3: ...Sensor Probe is clean and dry If it is not clean then gently wipe it off with a lint free swab 2 Mounting the Display Panel a Mount the Display Panel in a temperature humidity of less than 50 122 35 8...

Страница 4: ...E 01 and E 02 occurring Page 3 Mark Name Range Application C 01 Lower limited value of the UV low Alarm 0 100 According the actual requirement set the lamp low limited warming line In case the inspect...

Страница 5: ...t flashing bit is shifted d Press Page down Key flashing number will be reduce by one press Press Set Key to chose the digit then press Page down Key chose the number Fix it by press Set Key 3seconds...

Страница 6: ...Radiometer 6 a Connect AC power Terminal 9 and 10 and connect Sensor cable to terminal 5 Blue wire 3 Brown wire 4 Black wire Please refer to Fig 6 b Relay output terminal is 6 8 If you want to use rel...

Страница 7: ...abnormal Green light means normal Power cable Power cable Green light ballast Capacitor 220 2kV Page 6 In normal status the relay will connect 6 and 7 the Green light will be ON and Red light will be...

Страница 8: ...n 7 A S Request in Case of Product Failure 1 Should any failure is found in product please call the sales company or customer center for A S 2 Product warranty period is 1 year from the date of procur...

Отзывы: