GNF - 722 - 12 (0)                                                         Genicom Co., Ltd.                                                            A4(297

×

210

Page 2 

3. Production Installation 

  1) Mounting the Sensor Probe 

    a. Operation temperature of Sensor Probe is -30 ~ 85 

 (-22 ~ 185

    b. UV sensor of Sensor Probe and UV light source should be fixed in set distance to do verticality. 

       Once a mounting distance has been determined, make sure to use the same distance for any additional sensors used  

       for other lamps; doing so will provide a more accurate comparison of irradiance among different lamps.  

   c. In LW series, Sensor Probe connects after close teflon tape 3~4 times in screw page. 

   d. After mounting and positioning the sensor probe, make sure the window of Sensor Probe is clean and dry.  

       If it is not clean, then gently wipe it off with a lint free swab. 

 

 2) Mounting the Display Panel 

    a. Mount the Display Panel in a temperature & humidity of less than 50 

 (122

) & 35 ~ 85 %RH.   

    b. 3 wires of Sensor Probe connect to #3, #4 and #5 of Display. 

        [3 : +5V (Brown), 4 : GND (Black), 5 : Vin (Blue)].  

    c. Power cable connect to #9 and  #10 of Display.  (Power is AC voltage.  Be careful!)      

    d. Alarm for UV lamp running overtime #1 and #2 of Display. 

    e. Relay can use the 5A / 230VAC, and its criterion is the setting value of C-01.  

        #8 is N/O (Normal Open), #6 is N/C (Normal Close) and # 7 is Comm (Common).  

       If you want to get the short signal under 70 %, you should set 70 in C-01and connect the each wires  

       in #7 and #8 of Display. 

Fig. 2  Connection Diagram of UV Radiometer 6 

Fig. 3  Connection picture of UV Radiometer 6 

Sensor Probe 

AC Power 

9  

10 

3(Brown) 

4(Black) 

5(Blue) 

Содержание MG-06

Страница 1: ...and Applications 2 Product Specification and Configuration 3 Product Installation 4 Operation 5 Relay Operation method 6 Relative Reaction Curves of UV Sensors 7 A S Request in Case of Product Failur...

Страница 2: ...Configuration 1 Display Panel Panel Size 48 48 85 5 Panel cutting size 45 45 Tolerance 0 0 5 Power Supply 85 265 VAC 50 60 Consumption 3 W Operating Temperature 0 50 Operating Humidity 35 85 RH Relay...

Страница 3: ...Sensor Probe is clean and dry If it is not clean then gently wipe it off with a lint free swab 2 Mounting the Display Panel a Mount the Display Panel in a temperature humidity of less than 50 122 35 8...

Страница 4: ...E 01 and E 02 occurring Page 3 Mark Name Range Application C 01 Lower limited value of the UV low Alarm 0 100 According the actual requirement set the lamp low limited warming line In case the inspect...

Страница 5: ...t flashing bit is shifted d Press Page down Key flashing number will be reduce by one press Press Set Key to chose the digit then press Page down Key chose the number Fix it by press Set Key 3seconds...

Страница 6: ...Radiometer 6 a Connect AC power Terminal 9 and 10 and connect Sensor cable to terminal 5 Blue wire 3 Brown wire 4 Black wire Please refer to Fig 6 b Relay output terminal is 6 8 If you want to use rel...

Страница 7: ...abnormal Green light means normal Power cable Power cable Green light ballast Capacitor 220 2kV Page 6 In normal status the relay will connect 6 and 7 the Green light will be ON and Red light will be...

Страница 8: ...n 7 A S Request in Case of Product Failure 1 Should any failure is found in product please call the sales company or customer center for A S 2 Product warranty period is 1 year from the date of procur...

Отзывы: