Not enough equilibrium/destaining
solution in channel A. Causes and
Solutions
1. Not enough solution or the
opening of the tube is above
the solution. Please add more
corresponding solution.
2. There are twists in the tubes
which blocks solutions from
being pumped in. Please
untwist the tubes.
3. The ends of the tubes form a
seal with the bottom of the
container. Please shorten the
tube or make a slope cut at the
end of the tube.
4. Automatic adjustment due to
pressure changes. No special
actions required.
Press any key to return to the
main menu. Take the steps as
instructed.
Press to continue the
process.
If the red part in the display panel
is Channel B, press to
continue the process.
- 17 -
Problems
Solution
Description of the problem
Q1
Q2
Q3
starting up
(Q1~Q2)
During
Operation
Q3~Q6
The instrument detected there is
an incomplete program from the
latest working session. The
channels were forced to empty
for safety purposes.
Please wait for the instrument to
complete the self-check and
return to the Working Interface.
The accessory highlighted in the
blue part maybe Valve (1), Valve
(2), Valve(3), Channel A, or
Channel B
The lifetime of the part
highlighted in blue has reached
its recommended maximum
value.
Users are advised to contact
GenScript to replace the
corresponding components, so
as to avoid unnecessary losses
Return to working interface by
press any button.