par le LNE G'MED
2.
Recording patient info and starting the tests, two steps
2.1 En
ter patient’s info in
ROTAM software
In the « Patient record » tab,
enter the coordinates of the new
patient, or open his previous
created record
In the « preparation ROTAM
test
», under ‘settings’ rubric
select the tested knee and clarify
his status
enter the support-foot distance
(see §1.4)
2.2 Tighten the knee-cup and the thigh strap and launch the tests
Secure the knee-cup by tightening the 4 buckles to at least 60N (tighten in pairs, front/back).
The tightening must be infra-painful
Choose the rotation direction (intern or extern)
Launch the tests, at least twice, at 3, 5 and 8 Nm (if possible)
Note: the tightness is shown in real time under 'ROTAM status' and must be identical for a particular patient for
each test
Intern
Extern