15
1%-3% Virkon
TM
solution
4% glutaraldehyde solution
Chloramine T
Microcde SQTM 1:64
4% Decon
TM
90min
5.5.2 Decontamination Procedure
(1) Wear disposable gloves to protect yourself.
(2) Prepare the cleaning agent: 200ml 4% glutaraldehyde solution (or other reagents
recommended by the security officer).
(3) Empty the turntable.
(4) Turn off the power switch and disconnect the power cord.
(5) Use a cloth moistened with 70% ethanol to disinfect the outside of the instrument.
(6) Put the instrument in a big plastic bag. Make sure the front door is open.
(7) Put a piece of cloth soaked in glutaraldehyde solution in the plastic bag. Make sure
that the cloth does not touch the instrument.
(8) Seal the instrument in a plastic bag for at least 24 hours.
(9) Take the instrument out of the plastic bag.
(10) Clean the instrument with mild detergent.
(11) Use 70% ethanol to remove stains.
(12) After performing the decontamination procedure, attach a signed and dated
decontamination certificate on the transport package and attach it to the outside of the
package (see Appendix A: "Certificate of Decontamination").
5.6 Packaging for repair
If you need repair the instrument, Please package it following the guidelines below.
Caution
Before removing the instrument from the laboratory or performing any repair on
it, it must be thoroughly cleaned. When sending the instrument for repair, please remember:
(1)
Inform the reason for repair.
(2)
Decontaminate the instrument before hand. Pack the instrument according to its state before
disassembly.
(3)
Use the original packaging to ensure that the instrument will not be damaged during
transportation. Any damage will incur additional service charges.
(4)
Return the instrument (or other items) with a signed and dated decontamination certificate (see
Appendix A: "Certificate of Decontamination"). And attach it to the outside of the package.
(5)
After you contact the local agent or the manufacturer's technical service department, please
specify the disfunction.
(6)
Please refer to “Chapter 6” for more information on storage and transportation temperatures.
5.7 Service Contract
It is recommended that a service engineer trained by the manufacturer conduct a regular
maintenance and repair of the instrument every 12 months. This ensures that the product is properly
maintained and used without any problems.
5.8 Fill in the system log
System logs, including operation summary, maintenance procedures, error messages and other
useful information about system. These is very useful for proper maintenance of the system. Please
refer to Appendix B: "System Log". You can copy the form as many times as you need, but keep the
original blank form in the user manual.
5.9 Disposal of the instrument
If Purifier HT must be discarded, please follow the specific regulations of the laboratory
and the country for the treatment of biological waste. The disposal of the instrument is carried
out in accordance with the laws and regulations of the local authorities on the recycle of