12
4.2.4 User Settings
The user can select "
User Setting
" on the main screen to set the parameters of Purifier HT. After
entering the user setting interface, the user can set the language, fan switch, light switch and prompt
tone switch. There are two modes of language: Chinese and English. "
Factory Setting
" is not
accessible to the users. if necessary, please to contact the manufacturer. If the user uses 24 magnetic
rods, you can click "
P96 change to P24
" to switch to the 24-channel program.
4.2.5 UV Light Settings
Click "
UV Sterilization
" on the main interface to set UV disinfection, and set the time in the time
setting column. The remaining time is displayed as the time countdown. Click "
Start
" to start UV
disinfection. Click "
Close
" to stop UV disinfection.
4.3 Shutting Down
Turn off Purifier HT
:
(1) Press the power switch (Figure 2-2) on the rear panel of the instrument to the OFF position to
turn off Purifier HT. It is recommended to turn off the instrument at night and on weekends.
(2) Wipe the surface of the instrument with a soft cloth or paper towel with distilled water, mild
detergent (SDS, sodium lauryl sulfate) or soap solution.
(3) If contaminated reagents have been spilled on the workbench, please clean them with 70%
ethanol or other disinfectants (see "Cleaning Procedures").
Warning
Remove the deep-well plate and magnetic tip comb that are used in the
instrument. Dispose of all deep-well plates and magnetic tip comb as biohazard waste.