12

If the primary readout shows a value other than 

0.0%

, you must

calibrate the meter to operate in 

MI

mode. To do so, continue to

hold the meter in your hand (touching nothing) and press the 

ZERO

button. Immediately after the term 

ZERO

appears briefly below the

bar graph at the left of the display, the primary readout will show

0.0%

and the meter will be calibrated and ready to operate in 

MI

mode. 

To obtain accurate readings in 

MI

mode, you must recalibrate the meter each time you re-enter

that mode from another mode.

CHECKING CALIBRATION

Although you cannot calibrate the RHMG700DL yourself, you can 

check

the meter’s calibration

periodically to assure yourself of the accuracy of humidity-related readings (RH, absolute
humidity, dew point and mixing ratio). Factory calibration should minimize the humidity sensor’s
drift for several years—well beyond the warranty period.

At a minimum, a calibration check should entail immersing the slotted structure protecting the
humidity sensor in two bottles containing saturated salts. One reference salt should produce a
reading of a 33%; the other should produce a reading of 75%.

Two salts of this type are available from General: the HR33LJ (33%) MgCl Calibration Salt and the
HR75LJ (75%) NaCl Calibration Salt.

Depending on the size and shape of the calibration salt bottle that you use, you may have to
remove the hanger tab (Fig. 1, Callout B) from the meter to provide sufficient clearance for the
bottle to fit snugly over the slotted structure protecting the humidity/temperature sensor (after
removing the dust cap). The hanger tab is removable for that reason.

DATA LOGGING

This section explains how to use front-panel buttons on the RHMG700DL to view and manually
change data logging settings and to clear logged data from the meter's memory.

The final section of this manual—

Using the Included Software

, beginning on p. 15—outlines how

to do the same tasks more quickly and easily by pressing soft buttons on the X-In-1 application
included with the meter. 

Viewing Data Logging Settings

To view your data logging settings

, press the 

LOG/SET

button. If

the 

mode selector does not appear above 

TM

, press the 

<DN

or

UP>

button to move the selector above 

TM

, at the left side of the

top line of the display. Once the selector is above 

TM

, the two-digit

date portion of the system clock setting (09:18, or September 18,
2012 in the upper screen shot at right) will be visible on the
secondary readout. 

Содержание RHMG700DL

Страница 1: ...14 IN 1 DATA LOGGING THERMO HYGROMETER WITH PIN PINLESS MOISTURE METER USER S MANUAL RHMG700DL Please read this manual carefully and thoroughly before using this product...

Страница 2: ...urchasing General Tools Instruments General s RHMG700DL 14 in 1 Data Logging Thermo Hygrometer with Pin Pinless Moisture Meter Please read this user s manual carefully and thoroughly before using the...

Страница 3: ...s capacitance which the meter measures by gauging its effect on an electric field that the meter generates each time it is operated in internal moisture mode In internal moisture mode the meter exploi...

Страница 4: ...using pinless sensor 6 absolute moisture level using included pin type probe 7 dew point 8 wet bulb temperature 9 surface temperature requires optional K type thermocouple probe 10 condensation temper...

Страница 5: ...e Internal or MI measurements In memory and data logging modes used to clear stored readings and data H UNIT button In measurement mode each press toggles between Imperial and metric measurement units...

Страница 6: ...logging length setting indicator Data log working group setting indicator Clear data logging setting indicator Default setting indicator Alarm limit indicators Data logging indicator Zero indicator P...

Страница 7: ...midity temperature sensor To power off the meter press and hold the HOLD button for at least 2 seconds The shutdown will be announced by two short beeps After powering off the meter replace the sensor...

Страница 8: ...lluminates whenever any button is pressed It will remain on for 30 seconds and then extinguish to extend battery life Holding readings Briefly pressing the HOLD button freezes both the primary and sec...

Страница 9: ...isplayed returns the indicator to the record number of the last pair of readings stored To exit Recall mode and resume making real time measurements press the STORE ENTER button To erase all stored re...

Страница 10: ...hange your next step must be to press the STORE ENTER button Doing so saves the value shown on the primary readout as the selected parameter s new LOW setpoint It also causes LOW to disappear and retu...

Страница 11: ...ale they exist only to provide an analog way to quickly track changes in relative moisture levels To use the included pin type probe to measure the moisture level of a material 1 Use the UP or DN butt...

Страница 12: ...of 75 Two salts of this type are available from General the HR33LJ 33 MgCl Calibration Salt and the HR75LJ 75 NaCl Calibration Salt Depending on the size and shape of the calibration salt bottle that...

Страница 13: ...decrement the Year value press the STORE ENTER button When the Year value is correct press the LOG SET button to save the setting and advance to the Month field which should now be flashing Use the UN...

Страница 14: ...button to increment the TML value or the STORE ENTER button to decrement it The default value of TML is 15 minutes The minimum possible value is 1 minute and the maximum possible value is 30 days It s...

Страница 15: ...mini disc supplied with the RHMG700DL is to display data logs captured and stored by the meter The program s secondary function is to make it easy to set the meter s system clock and to change its da...

Страница 16: ...d parameters you wish to display At the upper left side of the screen are circles you can check to select the units of key measurement parameters Below these circles are six boxes for manually enterin...

Страница 17: ...9 Pinless Moisture Measurement Depth 0 75 in 19mm Dew Point Range 76 to 140 F 60 to 60 C Atmospheric Pressure Range 30 to 120 kPA Vapor Pressure Range 0 to 20 0 kPA Sampling Rate 2X sec Memory Capaci...

Страница 18: ...if after examination the company determines it to be defective in material or workmanship The warranty period begins on the date of purchase You are encouraged to register your product online General...

Страница 19: ...ntact our Customer Service to obtain an RGA Return Goods Authorization number before forwarding the unit via prepaid freight to the attention of our Service Center at this address General Tools Instru...

Страница 20: ...20...

Страница 21: ...1 6100 FAX 212 431 6499 TOLL FREE 800 697 8665 e mail sales generaltools com www generaltools com RHMG700DL User s Manual Specifications subject to change without notice 2014 GENERAL TOOLS INSTRUMENTS...

Отзывы: