5.6 Setup Mode Selection Block Diagram
This section is provided to aid the operator in making selections during the Setup Mode.
It is recommended that the operator fill-in the selections in the proper blanks and then
use this page as a reference while programming the Model TA102A. The blocks shown
below indicate the order of options in the
Setup
Mode. To the right of each block is a
description of the choices that are available for that option. More information about
making each selection is provided in Section 5.4.
Enter the Password, if the Password is enabled.
Enter the Inhibit Mode, if desired.
ENTER
SELECTION
Set the Energized (
En
) / De-Energized (
dE
) Option
Set the Latching (
LA
) / Non-Latching (
n
L) Option
Set the
A2
alarm set point (from 10 to 60, in increments of 5)
Set the Energized (
En
) / De-Energized (
dE
) Option
Set the Latching (
LA
) / Non-Latching (
nL
) Option
Set the
A1
alarm set point (from 10 to 60, in increments of 5)
The A1 Set Point cannot be higher than the A2 Set Point
Set the calibration level, LEL (from 25 to 90, in increments of 1)
Set the Fault Activate (
Ac
) or not (
nA
) during Inhibit Mode
Display will indicate “
ct
” for 5 seconds
Set the ramp time for the Card Test Mode (3 or 10 seconds)
Set the Alarm outputs for Active (
Ac
) or not Active (
nA
)
Set the Password to be Disabled (
Pd
) or Enabled (
PE
)
If the Password is Enabled:
Set
the
password
digits
Left
Right
After all of the options have been selected, the TA102A will enter the Setup
Check Mode.
Password
Inhibit
Mode ?
A2 Alarm
Options
A1 Alarm
Options
Calibration
Level
Fault/Inhibit
Options
Card Test
Options
Setup
Check Mode