Model TA102A
24
5.4 Setup & Setup Check Modes
The
Setup Check
Mode allows the operator to view the selected options for the module
without allowing any changes to be made. Once this mode has been entered, the
module will automatically display each of the selected options for a short period of time
and then it will return to normal operation. The
Setup
Mode allows the operator to
change the operating parameters by making choices for selected options.
The
Setup Check
&
Setup
Modes display identical information with the following
exceptions:
•
The
Setup Check
Mode allows the user to view the operating parameters of the
Model TA102A, whereas the
Setup
Mode allows the user to change the operating
parameters of the Model TA102A.
•
Entering the optional Password is only available in the
Setup
Mode.
•
The
Inhibit
Mode may only be entered from the
Setup
Mode. If the Inhibit Mode is
entered, the A1 & A2 outputs will be inhibited until the
Mode/Select
switch is
pressed.
NOTE -
The
Setup
and
Setup Check
Modes cannot be entered if the unit is in alarm or
fault.
During the Setup Mode the operator will be allowed to select options.
The selection procedure is the same for most of the options. Pressing the
Mode/Select
Switch toggles the available choices. When the display has indicated a choice for five
consecutive seconds, without the operator pressing the
Mode/Select
Switch, the
Setup
routine will accept that selection and move on to the next option available.
NOTE -
Before entering the
Setup
Mode to make changes, the user should fill out the
form and become familiar with the block diagram, Section 5.6, of this manual. This will
aid the user during the selection process in the
Setup
Mode.
The Password, the A1 & A2 Alarm set points and the calibration level options offer the
operator more than two choices. While these options are being selected, pressing the
Mode/Select
Switch repeatedly will sequence the display to the next available choice
for that option.
To Enter the
Setup Check
Mode or the
Setup
Mode. Press and hold the Mode/Select
switch until the
SETUP
LED begins flashing (about twenty seconds). When the
SETUP
LED is flashing, release the
Mode/Select
switch to enter the
Setup Check
Mode
(Figure 47). Continuing to press and hold the Mode/Select switch until the
SETUP
LED
stops flashing (about five seconds more) will allow the operator to enter the Setup
Mode. When the
SETUP
LED stops flashing and stays on, release the
Mode/Select
switch and the unit will enter the Setup Mode (Figure 47).