Model 2280A
20
system for field calibration of H
2
S sensors. The bottle is filled with a hydrogen sulfide
(H
2
S) in air mixture and is available in 7 concentrations. The temperature limitation for
operation and storage is 0°F to +130°F (-18°C to +54°C). Make sure the Model 2280A
has had power applied for at least 24 hours to ensure that the metal oxide film of the
sensor has stabilized. Make sure the sensor is seeing clean air.
Figure 8: Portable Purge Calibrator
To Use the Portable Purge Calibrator
1. Place the calibration cup over the sensor.
2. Apply the gas by opening the ON/OFF valve on the cylinder.
3. Allow the sensor to see the gas for1 to 2 minutes.
NOTE:
Do not store the cylinder with the regulator fully engaged in the cylinder valve.