GM7702 User Manual
- 14 -
The indicator has been calibrated on the analog output when it leaves the factory, user no
need to calibrate the analog output. If the analog output is abnormal, user can calibrate the
analog output by himself (it is recommended to calibrate under the guidance of a professional).
Calibration method (use a multimeter to connect the analog interface):
※
Select calibration point and press ENTER. Contact to
multimeter and input the display value, finally press
ENTER to complete.
※
Analog calibration has 4 steps to complete.
Change option by pressing MODE key.
1. Working parameter
2. Calibration parameter
1. 4mA calibration
2. 12mA calibration
3. 20mA calibration
4. 24mA calibration
※
In parameter interface, pressing TARE and ENTER 3s
to enter analog calibration.