Gendex VisualiX
TM
eHD
18
User and Service Manual
Doc #4519 986 19041 - April 2005
Gendex VisualiX
TM
eHD
19
User and Service Manual
Doc #4519 986 19041 - April 2005
VisualiX is supplied with two USB cables, 2 m and 4.5 m long.
The USB cables have different connectors at each end.
The Type A connector is
The Type B connector is
connected to the PC
connected to the IME
The VisualiX system may be connected and disconnected while the computer is on,
unless the PC presents particular limitations.
For information on installation of software programmes, please refer to the soft-
ware manual supplied with the software installed.
Indicator status
Colour
IME status
None
Off or initialising
Green
Ready for image acquisition
Red
System is not functioning properly
= off
= on
IME Intermediate Electronics may be installed:
-
externally to the PC
-
wall-mounted or on the dental assembly
The Intermediate Electronics unit has a bracket for wall assembly. This requires a standard
USB cable up to 5 m long for connection with the USB port on the PC.
This maximum cable length may be exceeded only if suitable active replicators (HUBs)
are used.
Содержание VisualiX eHD
Страница 1: ...VisualiX eHD User and Service Manual Doc 4519 986 19041 April 2005 ...
Страница 6: ...Gendex VisualiXTM eHD 4 User and Service Manual Doc 4519 986 19041 April 2005 ...
Страница 30: ...Gendex VisualiXTM eHD 28 User and Service Manual Doc 4519 986 19041 April 2005 ...
Страница 34: ...Gendex VisualiXTM eHD 32 User and Service Manual Doc 4519 986 19041 April 2005 ...
Страница 37: ...Gendex VisualiXTM eHD 35 User and Service Manual Doc 4519 986 19041 April 2005 ...