Appendix A. Technical Information 

GXDP-300

Leakage technique factors

PANORAMIC
 

4800 mAs/h, exposure with maximum values

 

(73 kV, 8 mA, 12 s) according to the 1:5 duty cycle 

Measurement bases

The kV is measured by monitoring differentially the current fl owing through 450 
Mohm, 1% feedback resistor connected between the tube anode and ground.
The mA is measured by monitoring current in the HT return line, which equals the 
tube current.

Collimator

TYPE
 BLD-M-1
PRIMARY SLIT
 

Adult panoramic slit only. For child panoramic the exposure time is reduced  

 

to give a reduced length image. 

PRIMARY SLIT SIZE 
 

0.7 - 0.75 x 38 mm

Z-motor 

DUTY-CYCLE
 

-Intermediate use: 6.25%, 25s ON, 400s OFF

Environmental data

OPERATING
 

- Ambient temperature from +10ºC to +40ºC

 

- Relative humidity 10 - 90%, no condensation

STORAGE/TRANSPORTATION
 

- Ambient temperature from -20ºC to +50ºC

 

- Relative humidity 5 - 85% no condensation 

 

- Atmospheric pressure 500 - 1080 mbar

A-4 

 

 

 

 

 

 

 

 

            

     

 

 

 

 

 

 

         

Appendix A. Technical Information 

GXDP-300

Leakage technique factors

PANORAMIC
 

4800 mAs/h, exposure with maximum values

 

(73 kV, 8 mA, 12 s) according to the 1:5 duty cycle 

Measurement bases

The kV is measured by monitoring differentially the current fl owing through 450 
Mohm, 1% feedback resistor connected between the tube anode and ground.
The mA is measured by monitoring current in the HT return line, which equals the 
tube current.

Collimator

TYPE
 BLD-M-1
PRIMARY SLIT
 

Adult panoramic slit only. For child panoramic the exposure time is reduced  

 

to give a reduced length image. 

PRIMARY SLIT SIZE 
 

0.7 - 0.75 x 38 mm

Z-motor 

DUTY-CYCLE
 

-Intermediate use: 6.25%, 25s ON, 400s OFF

Environmental data

OPERATING
 

- Ambient temperature from +10ºC to +40ºC

 

- Relative humidity 10 - 90%, no condensation

STORAGE/TRANSPORTATION
 

- Ambient temperature from -20ºC to +50ºC

 

- Relative humidity 5 - 85% no condensation 

 

- Atmospheric pressure 500 - 1080 mbar

Reviewed: Westling Lauri 22.06.11 11:27:45
Approved: Weckman Janne 22.06.11 13:46:25
See the PaloDEx Group Oy PDM system to determine the status of this document. Printed out: 05.09.11 15:25:27

D506222

Approved

Содержание GXDP-300

Страница 1: ...r Manual Digital Panoramic X ray Unit REF DP300UMEW ver 1 206259 rev 1 Reviewed Westling Lauri 22 06 11 11 27 45 Approved Weckman Janne 22 06 11 13 46 25 See the PaloDEx Group Oy PDM system to determi...

Страница 2: ...iewed Westling Lauri 22 06 11 11 27 45 Approved Weckman Janne 22 06 11 13 46 25 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 05 09 11 15 25 27 D506222 Appro...

Страница 3: ...umentation for Gendex GXDP 300TM and the operating software are copyrighted with all rights reserved Under the copyright laws this documentation may not be copied photocopied reproduced translated or...

Страница 4: ...iewed Westling Lauri 22 06 11 11 27 45 Approved Weckman Janne 22 06 11 13 46 25 See the PaloDEx Group Oy PDM system to determine the status of this document Printed out 05 09 11 15 25 27 D506222 Appro...

Страница 5: ...P 300 Contents User Manual v Contents 1 Introduction 1 1 1 GXDP 300 X ray unit 1 1 2 About this manual 1 1 3 Setup properties 1 2 Unit description 2 2 1 Main parts 2 2 2 Unit controls 4 2 3 Accessorie...

Страница 6: ...tents GXDP 300 vi User Manual Reviewed Westling Lauri 22 06 11 11 27 45 Approved Weckman Janne 22 06 11 13 46 25 See the PaloDEx Group Oy PDM system to determine the status of this document Printed ou...

Страница 7: ...r Manual 1 GXDP 300 1 Introduction 1 Introduction 1 1 GXDP 300 X ray unit The GXDP 300 the unit is a digital panoramic dental x ray unit designed to take panoramic exposures of large medium and small...

Страница 8: ...ser Manual 2 Unit description GXDP 300 2 Unit description 2 1 Main parts 1 Column 2 Motorized carriage 3 Rotating unit 4 Exposure warning lights 5 Emergency stop button Press to stop rotate to release...

Страница 9: ...r Manual 3 GXDP 300 2 Unit description 1 Head support 2 Midsagittal laser 3 Mirror 4 Frankfort laser and laser positioning knob 5 Cuspid positioning knob 6 Chin rest 7 Cuspid laser 8 Patient handgrips...

Страница 10: ...ser Manual 2 Unit description GXDP 300 2 2 Unit controls A Side control panel 1 Lasers key switches the patient positioning lasers on and off 2 Up key drives the unit up 3 Down key drives the unit dow...

Страница 11: ...r Manual 5 GXDP 300 2 Unit description 2 3 Accessories Chin rest Bite guide Bite rod 71 mm Chin support Partial edentulous bite positioner Nose support long for small adults and children Nose support...

Страница 12: ...ser Manual 3 Taking an Exposure GXDP 300 3 Using the Unit IMPORTANT NOTE If the unit is being used for the first time or if you are using the unit for the first time check that it is set up to your re...

Страница 13: ...P 300 3 Taking an Exposure User Manual 7 3 2 Taking Exposures Panoramic adult child and bitewing 1 Slide the chin rest on to the support holder 2 Dentate patients Attach the bite guide to the bite rod...

Страница 14: ...ser Manual 3 Taking an Exposure GXDP 300 3 Place the appropriate hygienic barrier on to the patient support you are using 4 Select the panoramic program you require panoramic Pan or bitewing B W The m...

Страница 15: ...P 300 3 Taking an Exposure User Manual 9 7 Press the Up Down keys to adjust the height of the chin rest so it is slightly higher than the pa tient s chin so that the patient will have to stretch up to...

Страница 16: ...User Manual 3 Taking an Exposure GXDP 300 10 Look at the reflection of the patient in the mir ror and position the midsagittal plane of the patient so that it coincides with the midsagittal plane lase...

Страница 17: ...P 300 3 Taking an Exposure User Manual 11 CAUTION When pressing the Up Down keys to adjust the tilt of the patient s head take care not to cause the patient any distress or discomfort 12 The cuspid la...

Страница 18: ...User Manual 3 Taking an Exposure GXDP 300 13 Carefully push the head support towards the patient A Make sure that patient s neck is stretched and straight Rotate the head support knob B clockwise to c...

Страница 19: ...P 300 3 Taking an Exposure User Manual 13 17 Press and hold down the exposure button for the duration of the exposure During the ex posure you will hear an audible signal and the exposure warning ligh...

Страница 20: ...User Manual 3 Taking an Exposure GXDP 300 20 Press the Return key to drive the unit to the patient entry position NOTE After the exposure a timer indicating the tube head cooling time will appear on t...

Страница 21: ...P 300 3 Taking an Exposure User Manual 15 Temporomandibular Joint TMJ 1 Slide the nose support into the support holder Use the short version for large and medium adults and the long version for small...

Страница 22: ...User Manual 3 Taking an Exposure GXDP 300 6 Ask the patient to remove any eye glasses false teeth jewellery and hair clips and pins Place a protective lead apron over the patient s shoul ders NOTE If...

Страница 23: ...P 300 3 Taking an Exposure User Manual 17 9 Press the Lasers key to switch the patient po sitioning lasers on They will remain on for two minutes NOTE The patient positioning lasers will automatically...

Страница 24: ...User Manual 3 Taking an Exposure GXDP 300 12 Carefully push the head support towards the patient A Make sure that patient s neck is stretched and straight Rotate the head support knob B clockwise to c...

Страница 25: ...P 300 3 Taking an Exposure User Manual 19 16 Press and hold down the exposure button for the duration of the exposure During the ex posure you will hear an audible signal and the exposure warning ligh...

Страница 26: ...User Manual 3 Taking an Exposure GXDP 300 19 Turn the head support knob B counterclock wise to open the temple supports Press the release button C at the top of the head support and then slide the hea...

Страница 27: ...P 300 4 Operating the unit without x rays User Manual 21 4 Operating the unit without x rays In some situations for example with nervous pa tients or patients with unusual anatomy you may wish to oper...

Страница 28: ...User Manual 5 Troubleshooting and Maintenance GXDP 300 5 Troubleshooting and Maintenance 5 1 Error messages and symbols If the unit is not used correctly or the unit malfunc tions an error message or...

Страница 29: ...r Manual 23 GXDP 300 5 Troubleshooting and Maintenance Error symbols REASON i The PC connected to the unit is not on ii GxPicture in the PC is not open iii The cable connecting the unit to the PC is d...

Страница 30: ...User Manual 5 Troubleshooting and Maintenance GXDP 300 System errors E4 REASON Tubehead too hot or too cold SOLUTION When the error message automatically clears the tubehead has reached the cor rect o...

Страница 31: ...r Manual 25 GXDP 300 5 Troubleshooting and Maintenance 5 2 Care and Maintenance Cleaning and disinfecting the unit Warning Switch the unit off before cleaning it Surfaces that the patient touches All...

Страница 32: ...User Manual 5 Troubleshooting and Maintenance GXDP 300 Correct operation of the unit If any of the unit s controls displays or functions fail to operate or do not operate in the way described in this...

Страница 33: ...r Manual 27 GXDP 300 6 Warnings and precautions 6 Warnings and precautions 6 1 General warnings The unit must only be used to take the dental x ray exposures described in this manual The unit must NOT...

Страница 34: ...User Manual 6 Warnings and precautions GXDP 300 6 2 User patient warnings The unit may be dangerous to the user and the patient if the safety regulations in this manual are ignored if the unit is not...

Страница 35: ...r Manual 29 GXDP 300 6 Warnings and precautions Disinfect all the surfaces that the patient is in contact with after every patient If the unit does not appear to be working cor rectly switch the unit...

Страница 36: ...User Manual 7 Disposal GXDP 300 7 Disposal At the end of useful service life of the device its spare parts its replacement parts and its acces sories make sure that you follow all local national and i...

Страница 37: ...r Manual 31 GXDP 300 7 Disposal Reviewed Westling Lauri 22 06 11 11 27 45 Approved Weckman Janne 22 06 11 13 46 25 See the PaloDEx Group Oy PDM system to determine the status of this document Printed...

Страница 38: ...GXDP 300 Appendix A Technical Information Appendix A Technical Information A 1 Technical specifications Type GXDP 300 Classification Complies with IEC 60601 1 1995 IEC 60601 2 7 1998 IEC 60601 2 28 1...

Страница 39: ...Appendix A Technical Information GXDP 300 TARGET MATERIAL Tungsten NOMINAL X RAY TUBE VOLTAGE 66 70 and 73 kV 4 kV 8 mA NOMINAL X RAY TUBE CURRENT 8 mA 1 mA 73 kV NOMINAL ANODE INPUT POWER 584 W nomi...

Страница 40: ...GXDP 300 Appendix A Technical Information LINE SAFETY SWITCH when required 100 120 VAC Approved type min 16 A 250 VAC 220 240 VAC Approved type min 10 A 250 VAC EARTH LEAKAGE CIRCUIT BREAKER when req...

Страница 41: ...Appendix A Technical Information GXDP 300 Leakage technique factors PANORAMIC 4800 mAs h exposure with maximum values 73 kV 8 mA 12 s according to the 1 5 duty cycle Measurement bases The kV is measu...

Страница 42: ...GXDP 300 Appendix A Technical Information System requirements and connections The PC and any other external device s connected to the system must meet the IEC 60950 standard minimum requirements Devi...

Страница 43: ...Appendix A Technical Information GXDP 300 Tube housing assembly cooling heating characteristics Tube rating chart Toshiba D 052 SB Reviewed Westling Lauri 22 06 11 11 27 45 Approved Weckman Janne 22...

Страница 44: ...GXDP 300 Appendix A Technical Information Anode thermal characteristics Reviewed Westling Lauri 22 06 11 11 27 45 Approved Weckman Janne 22 06 11 13 46 25 See the PaloDEx Group Oy PDM system to deter...

Страница 45: ...Appendix A Technical Information GXDP 300 A 2 Unit dimensions Reviewed Westling Lauri 22 06 11 11 27 45 Approved Weckman Janne 22 06 11 13 46 25 See the PaloDEx Group Oy PDM system to determine the s...

Страница 46: ...GXDP 300 Appendix A Technical Information A 3 Symbols that appear on the unit Reviewed Westling Lauri 22 06 11 11 27 45 Approved Weckman Janne 22 06 11 13 46 25 See the PaloDEx Group Oy PDM system to...

Страница 47: ...0 Appendix A Technical Information GXDP 300 Reviewed Westling Lauri 22 06 11 11 27 45 Approved Weckman Janne 22 06 11 13 46 25 See the PaloDEx Group Oy PDM system to determine the status of this docum...

Страница 48: ...1 GXDP 300 Appendix A Technical Information A 4 EMC declaration Guidance and manufacturer s declaration electromagnetic emissions The GXDP 300 is intended for use in the electromagnetic environment sp...

Страница 49: ...2 Appendix A Technical Information GXDP 300 Guidance and manufacturer s declaration electromagnetic immunity The GXDP 300 is intended for use in the electromagnetic environment specified below The cus...

Страница 50: ...3 GXDP 300 Appendix A Technical Information Guidance and manufacturer s declaration electromagnetic immunity The GXDP 300 is intended for use in the electromagnetic environment specified below The cus...

Страница 51: ...4 Appendix A Technical Information GXDP 300 Recommended separation distances between portable and mobile RF communications equipment and the GXDP 300 The GXDP 300 is intended for use in an electromagn...

Отзывы: