3.1 %AANNTTCCFF
Name
:
Configuration
Description
:
Sets address, type code, baudrate, data format
Syntax
:
%AANNTTCCFF(cr)
% delimiter character.
AA address of setting module (00-FF)
NN New address (00-FF)
TT New type
CC New baudrate
FF New data format
When changing baudrate or checksum, we should INIT* termination land.
7
6 5 4 3 2 1 0
Not used
Figure 3-1 Data format setting of AI modules
Response
:
!AA(cr) if the command was valid.
?AA(cr) if an invalid operation was entered. If
the INIT* terminal was not grounded when attempting
to change baud rate or checksum settings.
Syntax error or communication error may get no response.
! command is valid.
0=60Hz Restrain 1=50Hz Restrain
00=Engineer Unit Format
01=Percent Format
10=2’s Complement HEX Format
Checksum status
0
:
Disable
1
:
Enable