How to use Reading Scale for HCT
:
:
:
:
When Rotor No. F-2415 is used for the Hematocrit test of HCT value,
the following direction is the way to read the scale:
1
、
Remove the capillary tubes from the rotor head carefully
;
Warning:
Do not let the blood drain out of the tubes.
2
、
Place the capillary tubes onto grooves (#1) of tubes holder
carefully
;
let the extreme end (red blood cells) of capillary tubes
reach the straight line (#3) on right hand of the platform
;
3
、
Move the tubes holder (#2) to the position where the extreme
end of Leucocyte (#9) meets the diagonal line (#4) on platform
;
4
、
Move the sliding bar (#5) to meet the demarcation (#7) place
where separates the Leucocyte and red blood cells
;
5
、
The percentage of HCT is then read directly from the scale (#8)
at upper hand of the reading scale.
NOTE
:
:
:
:
-
Use clay to block the end of capillary tube which has mark on it
;
-
Refer laboratory method of reading scale to licensed medical
personnel.