When the device is turned on

, this control system carries out a complete self-test cycle over the hardware and 

the software (all the Led and displays light up, and the loudspeaker is activated). If everything works properly, this 
phase ends by a short acoustic signal.  

 

If the system detects any problem, the faults are indicated as described above, and the operator must either try to 
eliminate all the possible causes, or switch the equipment off and ask for a technical assistance.  

When the device is running

, this control system checks the functioning of the unit, included the output powers 

(2000 times per second), by repeating the complete test on the functioning every 20 minutes (in an activation 
interval). 

Even in this case, the problems are indicated as described above 

(“

When the device is turned on

”). 

Control of the Error Codes  

The unit stores the last 32 Error Codes that have been detected, in order to help the operators in finding a 
solution for the problems/faults. The procedure to follow for this control is detailed into the Service Manual.  

Inside  the  E

2

PROM  hexadecimal  value  corresponding  to  one  of  error  codes  reported on “Error  code”  column in  the 

above table are stored. 

Error description 

Error code 

Stored code in 

E

2

PROM (select “0d” 

in CAL mode)

1

 

Codes for Wrong Use or Activation Circuits Fault 

For all the “

U” 

problems, the operators can act by eliminating the cause of the intervention of the self-diagnosis system. 

 In all the other cases, ask for the technical assistance, after having checked the signal by switching off and on the unit  

U)

   Problem into the neutral electrode (broken cable, non connected cable to the unit or to 

 

the neutral electrode, neutral electr

ode “split” type not attached to the tissues 

no nP 

Failure into the neutral electrode control circuit (internal fault) 

Err nPC 

81 

Failure into the control circuit of the contact between the neutral electrode/the patient 
(internal fault)  

Err 53 

35 

 

Failure of an activation hand-switch at the switching on 

U)

   Activation hand-switch pressed at the switching on 

Err Hnd 

82 

 

Failure of a pedal foot-switch at the switching on 

U)

   Pedal foot-switch pressed at the switching on  

Err PEd 

83 

U)

   Contemporary activation not allowed of the two activation devices 

USr Act 

U)

   Key on the control panel pressed at the switching on  

 

Failure of a key on the control panel 

Err 14 

0E 

Codes for System Fault

 

For this kind of problems, ask for the technical assistance, after having checked the signal by switching off and on the unit 

E

RRORS RELATED TO 

M

ASTER MICROCONTROLLER

 

RAM Memory 

Err 32 

20 

FLASH Memory 

Err 34 

22 

System complemented variables 

Err 35 

23 

Watchdog Timer 

Err 33 

21 

Errors related to power supply section 

+5VDC failure 

Err 36 

24 

HF power supply failure with lower setting value 

Err 37 

25 

HF power supply failure with higher setting value 

Err 38 

26 

HF power supply higher than setting value during activation  

Err 97 

61 

Errors related to serial communication 

Failure regarding serial communication peripheral placed on mother board alsa code 
801463 

2 slow sound 

repeatitions; after 

fast repeatitions 

3d 

Failure regarding serial communication peripheral placed on power supply board alsa 
code 801471 

3 slow sound 

repeatitions; after 

fast repeatitions 

3E 

Failure regarding serial communication peripheral placed on double handswitch & nP 
control board alsa code 801462 

4 slow sound 

repeatitions; after 

fast repeatitions 

3F 

E

2

PROM Communication error 

5 slow sound 

repeatitions; after 

fast repeatitions 

40 

                                                        

 

Содержание ESU-X 300 NT

Страница 1: ...stra e 2 78532 Tuttlingen Germany Tel 49 7461 96624 0 FAX 49 7461 96624 22 E Mail info geister com Read instruction for use 0123 Medical product higher class I and shortcut for Notified Body T V S D P...

Страница 2: ...oagulation G Alarm Led for the HF leakage currents control circuit H Setting of the standard twin foot switch pedal functioning PEDAL 1 I Setting of the connection sockets for the bipolar electrodes l...

Страница 3: ...th sufficient density the cellular liquid of the tissues it warms it and generates what follows 1 a heating which is so rapid that the vapour pressure into the cells breaks their membranes and provoke...

Страница 4: ...an electrode in bad conditions or an electrode with connection defects causes a reduction of the output power as it does not have any good contact with the tissues 6 Remember that even if compliant wi...

Страница 5: ...circuit because these currents represent one possible source for undesired burns on the patient or the operators i e a patient who gets in touch with a metallic part of the operating table or with wet...

Страница 6: ...dal or coagulation blue pedal BIP to activate the bipolar currents of cut coagulating cut yellow pedal or coagulation blue pedal except for the coagulation AUTO which is not activated by any pedal foo...

Страница 7: ...d switch handle forceps for coagulation with hand switch for coag only in alternative Socket 5 A the led A under is lighted on handle forceps for coagulation laparoscopic instrument for the use by foo...

Страница 8: ...ust the power with the regulation keys UP DOWN For the delivery loud acoustic signal and yellow Led press on the yellow pedal of the twin foot switch or the yellow key of the hand switch handle Instru...

Страница 9: ...the delivery acute acoustic signal and blue Led press on the blue pedal of the twin foot switch or the blue key of the hand switch electrodes holder handle It is also possible to deliver the current b...

Страница 10: ...r the delivery acute acoustic signal and blue Led press on the blue pedal of the twin foot switch or the blue key of the hand switch electrodes holder handle It is also possible to deliver the current...

Страница 11: ...Dateiname mit Revisionsstand erstellt am erstellt von gepr ft von freigegeben von Seite ESU X 300NT_EN B 18 04 2018 ssta cgro cgro 11 von 35...

Страница 12: ...USE OF THE PEDAL FOOT SWITCHES Press on the yellow pedal for the pure coagulating cut currents and the blue pedal for the coagulation currents C BIPOLAR FUNCTIONING The bipolar functioning is always...

Страница 13: ...ith the active electrode This current is controlled by an automatic system ADC System for the self regulation of the functioning which guarantees that the power selected by the operators remains const...

Страница 14: ...med with forceps surgical instrument and for coagulations performed directly with the active electrode even far away from the tissues This current is controlled by an automatic system ADC System for t...

Страница 15: ...ted above the unit allows to A Use the resectoscope by foot switch activation See PAR CONNECTION AND USE OF THE PEDAL FOOT SWITCHES Just press on the yellow pedal for the pure cutting coagulating cutt...

Страница 16: ...keys UP DOWN For the delivery acute acoustic signal and blue Led press on the blue pedal of the twin foot switch Instructions for use adjustments and electrodes Use the loop ball electrodes for the r...

Страница 17: ...for a cut without coagulating effect Current BLEND 1 from 90 100W on the prostate for a cut with soft coagulating effect For the Cut with ribbon loop electrodes use as follows Current PURE from 160 18...

Страница 18: ...informs the operators that the neutral electrode is not connected Impossibility to use the monopolar currents the displays for the monopolar currents powers are switched off Activation modes and outp...

Страница 19: ...he unit 0 5sec on the left 5sec on the right For the switching off it is possible either to disconnect the forceps from the tissues or wait until the current disconnects automatically once the tissues...

Страница 20: ...thout any sticking of the tissues on the electrode the above mentioned powers maybe have to be changed through small variations of 10 15W each time For the Coagulation When bipolar forceps are used it...

Страница 21: ...ion Is it possible to modify the adjustments of the power s and or the selected settings of a memory before the use Yes it is always possible through the different adjustment selection keys and this v...

Страница 22: ...gon gas flow if it is used can obviously be changed by setting the desired levels and by pressing the memorisation key MEM of the area B The pre setting of these memories allows to use the unit as fol...

Страница 23: ...raph PRE SET MEMORIES nr 8 Use in flexible endoscopic surgery or in endoscopic surgery under liquid The pre setting concerns 1 Activation mode and output available 2 In these programs are not set outp...

Страница 24: ...o the neutral electrode broken cable non connected cable to the unit or to the neutral electrode neutral electrode split type not attached to the tissues no nP Failure into the neutral electrode contr...

Страница 25: ...nopolar SPRAY mode Err 46 2E Output peak voltage reading failed lower than expected Err 47 2F Output peak voltage reading failed higher than expected Err 48 30 Output peak current reading failed lower...

Страница 26: ...ins Fuses T 5 A Monopolar mode with the possibility to connect 1 or 2 electrodes holder handles both of them with hand switches or just one with hand switches and the other one with pedal foot switche...

Страница 27: ...Directive 89 336 CEE Category A Suggested distances to keep from not vital devices Source of the Current RF Typical Power W Distance m Microcellular telephones CT1 CT2 CT3 0 01 0 4 Mobile telephones D...

Страница 28: ...300NT_EN B 18 04 2018 ssta cgro cgro 28 von 35 MONOPOLAR CURRENTS Variation of the output power from 50 to 2000 by selecting 100 and 50 of maximum power Increase in the output power with nominal load...

Страница 29: ...50 200 250 300 3500 3000 2500 2000 1000 500 1500 20 40 60 80 100 120 140 7000 6000 5000 4000 3000 2000 1000 7500 Vpp V 50 100 50 100 50 100 P W RMS Impostazione Reglage P W RMS Setting Impostazione Re...

Страница 30: ...120 150 5000 4000 3000 2000 1000 50 100 150 200 250 3500 3000 2500 2000 1500 1000 500 50 100 50 100 50 100 200 300 400 500 600 700 800 900 1k 1 1k 1 2k 1 3k 1 4k 1 5k 1 6k 1 7k 1 8k 1 9k 2k 100 200 3...

Страница 31: ...0 200 240 280 3500 3000 2500 2000 1500 1000 500 20 40 60 80 100 120 140 7000 6000 5000 4000 3000 2000 1000 7500 50 100 50 100 200 300 400 500 600 700 800 900 1k 1 1k 1 2k 1 3k 1 4k 1 5k 1 6k 1 7k 1 8k...

Страница 32: ...X 300NT_EN B 18 04 2018 ssta cgro cgro 32 von 35 BIPOLAR CURRENTS Variation of the output power from 10 to 1000 by selecting 100 and 50 of maximum power Increase in the output power with nominal load...

Страница 33: ...40 20 130 120 100 80 60 40 20 130 120 100 80 60 40 20 20 40 60 80 100 120 140 160 20 40 60 80 100 120 130 20 40 60 80 100 120 130 40 80 120 160 20 60 100 140 100 200 300 400 500 600 700 800 850 20 40...

Страница 34: ...20 130 120 100 80 60 40 20 130 120 100 80 60 40 20 130 120 100 80 60 40 20 20 40 60 80 100 120 130 20 40 60 80 100 120 130 20 40 60 80 100 120 130 450 400 350 300 250 200 150 100 50 20 40 60 80 100 12...

Страница 35: ...RIES see next page for details Nominal voltage 4000 V E 1 Straight knife electrode E 5 Straight needle electrode thick type 2 pcs E 7 Straight needle electrode thin type E12 Straight ball electrode 2...

Отзывы: