Moisture Meter G650
i
I
27
This function shall be only used in case of occurring some fault in the download of Internet
curves of G650
i
. In this function, the standard equations built into the G650
i
will be uploaded,
and they may not be the most updated. Moreover, there is no way to select them, so all the
equations will be uploaded.
The equations undergo updates with time, and we strongly recommend that an Internet
connection be made in order to download the most updated curves.
To upload standard equations:
1.
Connect the G650
i
;
2.
Go to “Configuration” and depress YES; do not forget to put the Hard Lock in the
serial output;
3.
Go to “5.8 Uploading standard equation” function, depress “YES”, a new message
will show up indicating the upload of products in the memory:
4.
The system automatically completes this function.
Programing...
Product: 65
5.8. Uploading standard equation
How to exit “5.8 Uploading Internet Equation” function
This function passes the control of the G650
i
processor to the computer to which it is
connected. Once having entered in this function, we have only two alternatives to exit:
A.
Exit with the site commands;
B.
Disconnecting, waiting for 5 seconds, and connecting again the power
supply to G650
i
.
Содержание G650i
Страница 1: ...User manual Moisture Meter G650i Version 1 00 000 Date 06 20 2014...
Страница 2: ......
Страница 42: ...42 I MoistureMeterG650i Technical specifications...
Страница 43: ...MoistureMeterG650i I 43 Standard Products in this version and operating ranges for each Product...
Страница 46: ...Annotations...