Description of the signal inputs/outputs
32
Function
Test
time
Action
IN1
- Measurement limited to
parameter 1 or
measuring point 1
Fixed,
10
sec-
onds
Selection "1 measuring
point":
Only parameter 1 is measured as
long as the contact is open or
closed.
Selection "2 measuring
point":
Only measuring point 1 with pa-
rameter 1 is measured as long as
the contact is open or closed.
In
, select
=>
=>
Select the type of contact
Press "ENTER" to confirm the selection
Function
Test
time
Action
IN2
Water meter input
None
Quantity recording for start-
ing an analysis
Function
Connection
Action
OUT
Programmable interface
0-20 mA or 4-20 mA
max. load
500 Ohms
OR
Programmable voltage
interface
0 - 10 V or 2 - 10 V
OR
Serial interface RS 232
Serial bus
(2-wire cable)
See description of inter-
face card RS 910
(See Technical info "T2000-
RS232")
The section "Interfaces" contains a detailed description
IN1
Terminals 24,25
▼▲
IN2
Terminals 26,27
OUT
Terminals ┴, 28,29