Section 3-5 - Connection of Auxiliary Devices
Section 3-5
Connection of Auxiliary Devices
NOTE:
Normally auxiliary devices and peripherals come pre-installed with the system.
below outlines hardware installation procedures described in the sub-sections.
NOTE:
For more detailed Safety Considerations when connecting auxiliary devices to the Voluson™ P8 /
Voluson™ P6 system, please review:
Chapter 1 - Auxiliary Devices Safety, on page 1-8
.
Table 3-7
Connection Procedures
Sub-section
Description
Page Number
Connecting the Black & White Printer
Connecting the DeskJet Color Printer
Connecting the Wireless Network Adapter
Connecting the USB Flash Memory Stick
Connecting the external USB Hard disk (Handydrive)
General Remarks and Hints when using external USB-Devices
WARNING
WARNING
!! WARNING:
After each installation, the leakage currents have to be measured according to
IEC 60601-1 respectively UL 60601-1.
CAUTION
!! CAUTION:
Please observe that some printers may not be medical devices! If the Bluetooth Printer and/or
Line Printers are not medical devices, they have to be located outside of the patient
environment and use separation device such as LAN/USB isolator or isolation transformer
(according to IEC 60601-1 / UL 60601-1).
2,5 m
1,5 m
1,5 m
1,5 m