1. Press CW (8.5). A scanning screen is displayed with a

Doppler cursor on the 2D mode image and a Doppler
spectrum in the lower part of the screen.

2. Use the trackball (9.1) to position the Doppler cursor

line over the area of interest.

Tissue Velocity Imaging (TVI)

1. While in 2D mode press TVI (8.6) on the control panel.
2. Use the trackball (9.1) (assigned function: Pos) to

position the ROI frame over the area to be examined.

3. Press Set. The instruction Size should be highlighted in 

the trackball status bar.

NOTE: If the trackball control Pointer is selected, press Set to 
be able to select between Position and Size controls.

4. Use the trackball to adjust the dimension of the ROI.

Quick Card

Vivid S60 / S70

Trackball and related buttons

Select keys (9.2):
1. Angle: Sets the cut-planes and crop planes to pre-

defined positions.

2. Pos: Adjusts the ROI position

Update/Menu Key (9.3):
In Freeze: displays a pop-up System menu.
In Live duplex mode (Doppler or M-mode): toggles
Live/Freeze between the 2D image and the spectrum
image.

Trackball Key (9.4):
1. Scroll: scroll through the progress bar to watch the

cine loop frame by frame

2. Angle/Pos
3. Ptr: mouse left key

Upper key (9.5):
Configurable as either a select key, a pointer, an Image
store button, or an Image cursor.
To set the function:
1. Press Utility/Config on the Touch panel
2. Select Imaging/Global

4D Control

1. Enter 4D by pressing the 4D button (10) on the

control panel. The volume acquisition is started.

2. Adjust the desired parameters on the touch

panel.

Multi D Control

1. Press Multi-D (10) on the control panel

NOTE: The Bi-plane screen is displayed.

2. Press Tri-plane or Bi-plane on the touch panel to

activate the corresponding mode.

Loudspeaker volume control

Rotate (11) to adjust the volume level of the integrated
speakers.

PW

Color

1. From an optimized 2D image, press Color (8.3).
2. Use the trackball (9.1) (assigned function: Pos) to

position the ROI frame over the area to be examined.

3. Press Set. The instruction Size should be highlighted

in the trackball status bar.

NOTE: If the trackball control Pointer is selected, press

Trackball to be able to select between Position and Size

controls.

Use the trackball to adjust the dimension of the ROI.

1. Press PW (8.4). A scanning screen is displayed with a

Doppler cursor on the 2D mode image and a Doppler
spectrum in the lower part of the screen.

2. Use the trackball (9.1) to position the sample volume

location over the area of interest.

3. In PW, adjust the Sample Volume.

NOTE: Sample Volume adjustment may affect the Scale,
Frame rate and LV rej. settings.
NOTE: Adjustment of the Sample volume may affect the PRF
(Nyquist limit) settings.

Содержание Vivid S60

Страница 1: ...selection 2D Quick Card Vivid S60 S70 GE Healthcare 1 2 3 4 5 6 7 4D Control Multi D Control Loudspeaker volume ctl Flex Zoom Clear Depth P1 Print capture Freeze 2D Freeze Measure Caliper Store M Mode Color PW CW Tissue Velocity Imaging Trackball Select keys Update Menu Key Trackball Key Upper Key VIVID S70 ONLY ...

Страница 2: ...sor type on the Touch panel 3 With the trackball 9 1 adjust the position of the cursor line 4 Activate the desired mode M Mode 2D Press 2D 8 1 on the control panel to access 2D mode 1 To access M Mode from any other scan mode press MM 8 2 on the control panel 2 Use the trackball 9 1 to position the cursor over the required area of the image 3 Press Freeze 17 4 Use the trackball to scroll through t...

Страница 3: ... an Image store button or an Image cursor To set the function 1 Press Utility Config on the Touch panel 2 Select Imaging Global 4D Control 1 Enter 4D by pressing the 4D button 10 on the control panel The volume acquisition is started 2 Adjust the desired parameters on the touch panel Multi D Control 1 Press Multi D 10 on the control panel NOTE The Bi plane screen is displayed 2 Press Tri plane or ...

Страница 4: ...ignment configuration P1 button configuration NOTE The P1 button on the Control panel can be configured to perform several actions e g Video Print Laser print DICOM storage etc 1 Press Utility Config on the Touch panel 2 Select the Connectivity category and Additional outputs subgroup 3 In Button field select P1 4 Select an output device in the Available output field and press the Right arrow butt...

Страница 5: ...Store 2 If desired select and adjust the cineloop to store using the cineloop controls on the touch panel 3 Press Store 19 to save the cineloop NOTE The system can be configured to perform either Retrospective storage while scanning press Store to store the last elapsed defined number of cycles seconds Prospective storage while scanning press Store to start storage of the forthcoming defined numbe...

Отзывы: