Chapter 4 - Functional Checks
4-21
P R E L I M I N A R Y
4-3-1-3
System SN Test
4-3-1-4
Speakers Tests
Table 4-7
System SN Test
Step
Task
Expected Result(s)
1
Boot-up the System in Regular Scanning Mode
2
Press
Config
3
Under the
Connectivity
tab, select
TCPIP
then
Computer Name
4
Make sure the computer name is System ID
5
Select
Advanced Settings
6
Type
Desktop
and press <
Enter
>
7
Select
My Computer
and press
Menu
8
From the drop-down menu, select
Properties
and then
select the
Computer Name
tab.
9
Make sure the System SN is indicated under the
computer description.
The computer description matches the system
SN.
Table 4-8
Speakers Test
Step
Task
Expected Result(s)
Back End Path:
1
Boot-up the System in Maintenance Mode
2
Press
Alt +Config
to start the diagnostic package
3
From the diagnostic test tree, select
Back End
>
PC Doctor
>
Interactive
>
Audio Test Tree
.
4
Click
Start
.
5
At the WAV window, press the
Left Channel
button
Make sure that the
left
internal speaker is working
Verify that there is clear sound from all speakers at all
frequencies. Note that
right
and
left
are as seen when
standing behind the unit.
6
At the WAV window, press the
Right Channel
button
Make sure that the
right
internal speaker is working
Verify that there is clear sound from all speakers at all
frequencies. Note that
right
and
left
are as seen when
standing behind the unit.
7
At the WAV window, press the
Center Channel
button
Make sure that the
Sub woofer
internal speaker is
working. Verify that there is clear sound from all
speakers at all frequencies. Note that
right
and
left
are
as seen when standing behind the unit.
RFI Path:
1
From the diagnostic test tree, select
Front End Test
>
RFI
. Then select
RFI Audio Test
and press
Start
.
On completion, the circuitry test will indicate Pass.
Содержание Vivid S5 N
Страница 2: ......
Страница 4: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ii P R E L I M I N A R Y ...
Страница 5: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iii P R E L I M I N A R Y ...
Страница 6: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL iv P R E L I M I N A R Y ...
Страница 7: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL v P R E L I M I N A R Y ...
Страница 8: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vi P R E L I M I N A R Y ...
Страница 9: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL vii P R E L I M I N A R Y ...
Страница 10: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL viii P R E L I M I N A R Y ...
Страница 11: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL ix P R E L I M I N A R Y ...
Страница 12: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL x P R E L I M I N A R Y ...
Страница 13: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xi P R E L I M I N A R Y ...
Страница 14: ...GE PART NUMBER FN091065 REVISION 2 VS5 N AND VS6 N SERVICE MANUAL xii P R E L I M I N A R Y ...
Страница 833: ......
Страница 834: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 NO 3191 HORTEN NORWAY www gehealthcare com ...