Chapter 8 Replacement Procedures
8-63
8-6-2
Connector Board Assy 3PP (FRU No. 5314411-2) This is a description on how to
remove and replace the Conn. Board Assy.
8-6-2-1
Tools
•
Common Phillips screwdrivers
8-6-2-2
Needed Manpower
•
1 person, 15 minutes
8-6-2-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-6-2-4
Removal Procedure
NOTE: CAUTION
An Electronic discharge may damage a component.turn OFF power and wear the
wrist strap before you remove circuit boards. do not un plug the power card to keep ground continuity.
Do not bend or flex the boards when mounting /dismounting each boards surface mount IC boards are
very susceptible to damage from flex/torque.
1) Remove Right cover. Refer
for details on how to remove right cover
2) Remove Left cover. Refer
for details on how to remove right cover
3) Remove Front cover. Refer
section Figure 8-27 on page 33
for details on how to remove right cover
4) Loosen four Screws to remove the connector board and remove the conn board from the system.
. To open the Hinge Door loosen knob screw (A) & (B)