4400-0108 enFlow Preventive Maintenance Record Rev. C Apr-2011 

Page 4 of 8 

wire is inserted into the tubing just be certain that the wire is in contact with the fluid.  

5.  Next, put a non-venting cap on the open end of the extension set. 

6.  Place the cartridge setup in the warmer. 

7.  Connect the ECG lead from the leakage tester to the wire inserted into the extension set 

tubing.  

8.  Perform the “ECG” lead leakage test. 

 

 

 

 

 

 

 
Functional Tests 
 

1.

 

enCheck Model 400 

 

Equipment 

enCheck (Includes “K” type probe.) 
Thermal thermocouple meter with ± 5 °C 

accuracy 

enFlow Controller/AC Power Pack

 

enFlow Warmer

 

Purpose 

The enCheck Tester was developed to quickly and reliably trigger the over-temperature alarm 
condition on the enFlow Warmer. Within seconds, the enCheck unit will heat the Warmer to an over-
temperature scenario causing the alarm to sound. Additionally, the enCheck is designed to verify the 
Warmer operation at the enFlow’s installation site. 

Procedure

 

A.

 

Normal Mode - Simulated Use Performance Testing 

When the enCheck is connected and running in the normal mode, the heat is generated from the 
warmer unit using the same technology as when a cartridge is installed. This mode allows for 
confirmation of the temperature output of the Warmer. 

1.

 

Plug the Controller/AC Power Pack into a hospital grade outlet. 

2.

 

Connect the enCheck to the Controller/AC Power Pack by inserting the male plug end of the 
enCheck Hubbell connector into the female receptacle on the front face of the Controller/AC 
Power Pack. Push it in so that the plug cover is tight against the receptacle. 

3.

 

Next, connect the Warmer to the enCheck by inserting the male plug end of the Warmer into 
the enCheck female receptacle. 

4.

 

Take the temperature probe connector on the enCheck and insert into a thermometer. Set 
thermometer to “K” type setting.  

5.

 

Insert the enCheck unit into the Warmer. The end of the unit is keyed similar to the Cartridge 
so it will only fit in the correct orientation. Close the covers. 

6.

 

Confirm the enCheck is set to the normal mode. 

7.

 

Move the MAINS power switch on the back of the Controller/AC Power Pack to the ON 
position. Wait for the thermometer to stabilize, 

 30 to 60 sec. assuming all equipment is 

close to 20 °C. 

8.

 

The temperature on the thermometer should be 40 °C ± 2 °C. 

 
B.

 

Overheat Mode – Alarm Test 

When operated in the overheat mode, the heat is generated from the contact plate on the underside 
of the  enCheck unit. The enCheck heats the Warmer to > 45 °C. This temperature range simulates an 
over-temperature situation demonstrating the functionality of the alarm (audible, LED’s, and display 
(Controller only).  

1.

 

Setup is the same as in the above steps 1-7 with the exception of step 4.  Skip step 4. (In the 
Overheat mode, it is only important to review the temperature of the Warmer, which is 

Current Normal 

Single Fault 

Earth leakage

 

5 mA

 

10 mA

 

Touch/chassis leakage

 

100 µA

 

500 µA

 

Patient leakage

 

100 µA

 

500 µA

 

From Table I. Leakage current limits (from IEC 60601-1).

 

Содержание Vital Signs enFlow

Страница 1: ...enFlow IV Fluid Blood Warming System Preventive Maintenance Record Serial No ________________ Vital Signs a GE Healthcare Company Please consult Operator s Manual for Equipment Symbol List ...

Страница 2: ...olor Confirm the display panel Controller only shows in yellow the conditional message not heating 2 Connect the Warmer without a Disposable Cartridge inside to the Controller AC Power Pack Confirm the beep signaling connection Confirm the display Controller only continues to show the conditional message not heating Confirm the Warmer power LED is flashing green Electrical Safety Tests 1 Ground wi...

Страница 3: ...following configurations Allowable values of continuous LEAKAGE CURRENTS in µA ENCLOSURE LEAKAGE CURRENT TYPE BF Normal polarity normal ground Normal Condition 100 Reverse polarity normal ground Single Fault Condition 500 Reverse polarity open ground Double Fault Condition 500 Normal polarity open ground Single Fault Condition 500 3 Leakage current of the Warmer to the saline in the IV line Equipm...

Страница 4: ...le plug end of the enCheck Hubbell connector into the female receptacle on the front face of the Controller AC Power Pack Push it in so that the plug cover is tight against the receptacle 3 Next connect the Warmer to the enCheck by inserting the male plug end of the Warmer into the enCheck female receptacle 4 Take the temperature probe connector on the enCheck and insert into a thermometer Set the...

Страница 5: ...sion pump capable of maintaining up to 200 mL min IV line set Water Bath Source of distilled water or normal fluid 0 5 L at 20 C 2 0 C 2 Extension sets 22 9 cm 9 in Thermometer capable of measuring 10 C to 60 C accurate to 0 1 C plus 2 external K type probes Timer Waste water container Alternative equipment Graduated cylinder 100 to 500 mL Purpose This test is performed to ascertain that the outpu...

Страница 6: ... C 10 It is recommended to repeat steps 1 9 for the flow rate of 60 20 mL min High Flows 11 Repeat steps 1 7 for the high flow rates of 125 20 mL min 175 20 mL min and 200 mL min However in place of steps 8 and 9 measure the rise in temperature of the output fluid over the input fluid value The rise should be 16 5 C 3 Alternate Method Alarm Test Equipment enFlow system Power source Infusion pump c...

Страница 7: ...e temperature LED on the Warmer flashes red indicating that the fluid has gone over temperature Customer Service E mail customerserviceVSD ge com Phone 973 956 5300 800 932 0760 Technical Support Email enflowtechsupport ge com Phone 781 862 7008 FAX 781 862 7018 Service Center Address Enginivity LLC 1 Militia Dr LLA Lexington MA 02421 GE Medical Systems SCS 283 rue de la Miniere 78530 BUC FRANCE T...

Страница 8: ...ack Serial No__________________ Warming System Location Identifier____________________ Date Test Pass Fail or Recorded Value Date Initials Pass Fail or Recorded Value Date Initials Pass Fail or Recorded Value Date Initials Pass Fail or Recorded Value Date Initials Inspections Temp readout display and LED s Electrical Safety enCheck Alarm enCheck Temperature Output Alarm Performance Comments ...

Отзывы: