enFlow

 

IV Fluid / Blood Warming System 
 

 

 
Preventive  
Maintenance 
Record 
 
 
Serial No. ________________ 

 

 

 

 

 
Vital Signs 
a GE Healthcare Company 

 
 

 

 

Please consult Operator’s Manual for Equipment Symbol List

 

Содержание Vital Signs enFlow

Страница 1: ...enFlow IV Fluid Blood Warming System Preventive Maintenance Record Serial No ________________ Vital Signs a GE Healthcare Company Please consult Operator s Manual for Equipment Symbol List ...

Страница 2: ...olor Confirm the display panel Controller only shows in yellow the conditional message not heating 2 Connect the Warmer without a Disposable Cartridge inside to the Controller AC Power Pack Confirm the beep signaling connection Confirm the display Controller only continues to show the conditional message not heating Confirm the Warmer power LED is flashing green Electrical Safety Tests 1 Ground wi...

Страница 3: ...following configurations Allowable values of continuous LEAKAGE CURRENTS in µA ENCLOSURE LEAKAGE CURRENT TYPE BF Normal polarity normal ground Normal Condition 100 Reverse polarity normal ground Single Fault Condition 500 Reverse polarity open ground Double Fault Condition 500 Normal polarity open ground Single Fault Condition 500 3 Leakage current of the Warmer to the saline in the IV line Equipm...

Страница 4: ...le plug end of the enCheck Hubbell connector into the female receptacle on the front face of the Controller AC Power Pack Push it in so that the plug cover is tight against the receptacle 3 Next connect the Warmer to the enCheck by inserting the male plug end of the Warmer into the enCheck female receptacle 4 Take the temperature probe connector on the enCheck and insert into a thermometer Set the...

Страница 5: ...sion pump capable of maintaining up to 200 mL min IV line set Water Bath Source of distilled water or normal fluid 0 5 L at 20 C 2 0 C 2 Extension sets 22 9 cm 9 in Thermometer capable of measuring 10 C to 60 C accurate to 0 1 C plus 2 external K type probes Timer Waste water container Alternative equipment Graduated cylinder 100 to 500 mL Purpose This test is performed to ascertain that the outpu...

Страница 6: ... C 10 It is recommended to repeat steps 1 9 for the flow rate of 60 20 mL min High Flows 11 Repeat steps 1 7 for the high flow rates of 125 20 mL min 175 20 mL min and 200 mL min However in place of steps 8 and 9 measure the rise in temperature of the output fluid over the input fluid value The rise should be 16 5 C 3 Alternate Method Alarm Test Equipment enFlow system Power source Infusion pump c...

Страница 7: ...e temperature LED on the Warmer flashes red indicating that the fluid has gone over temperature Customer Service E mail customerserviceVSD ge com Phone 973 956 5300 800 932 0760 Technical Support Email enflowtechsupport ge com Phone 781 862 7008 FAX 781 862 7018 Service Center Address Enginivity LLC 1 Militia Dr LLA Lexington MA 02421 GE Medical Systems SCS 283 rue de la Miniere 78530 BUC FRANCE T...

Страница 8: ...ack Serial No__________________ Warming System Location Identifier____________________ Date Test Pass Fail or Recorded Value Date Initials Pass Fail or Recorded Value Date Initials Pass Fail or Recorded Value Date Initials Pass Fail or Recorded Value Date Initials Inspections Temp readout display and LED s Electrical Safety enCheck Alarm enCheck Temperature Output Alarm Performance Comments ...