Lockout/Tagout (LOTO) requirements
Versana Balance – Basic Service Manual
1-23
5808768-100 English Rev.9
1-8 Lockout/Tagout (LOTO)
requirements
Follow Lockout/Tagout requirements by ensuring you are in total
control of the AC power plug at all times during the service
process.
To apply Lockout/Tagout (LOTO):
1. Plan and prepare for shutdown.
2. Shutdown
the
equipment.
3. Isolate
the
equipment.
4. Remove/disconnect the battery, if present.
5. Apply
Lockout/Tagout
Devices.
6. Control all stored and residual energy.
7. Verify
isolation.
All potentially hazardous stored or residual energy is relieved.
WARNING
Energy Control and Power Lockout for Versana Balance.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait for at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
Ultrasound System components may be energized.
Содержание Versana Balance
Страница 6: ...i 4 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Страница 7: ...Versana Balance Basic Service Manual System view i 5 5808768 100 English Rev 9 ...
Страница 8: ...i 6 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Страница 9: ...Versana Balance Basic Service Manual System view i 7 5808768 100 English Rev 9 ...
Страница 10: ...i 8 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Страница 11: ...Versana Balance Basic Service Manual System view i 9 5808768 100 English Rev 9 ...
Страница 12: ...i 10 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Страница 13: ...Versana Balance Basic Service Manual System view i 11 5808768 100 English Rev 9 ...
Страница 14: ...i 12 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Страница 15: ...Versana Balance Basic Service Manual System view i 13 5808768 100 English Rev 9 ...
Страница 16: ...i 14 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Страница 17: ...Versana Balance Basic Service Manual System view i 15 5808768 100 English Rev 9 ...
Страница 28: ...i 26 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Страница 58: ...Introduction 1 30 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Страница 270: ...Components and Functions Theory 5 6 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Страница 278: ...Service Adjustments 6 8 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Страница 330: ...Diagnostics Troubleshooting 7 52 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Страница 376: ...Replacement Procedures 8 46 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Страница 390: ...Renewal Parts 9 14 System view Versana Balance Basic Service Manual 5808768 100 English Rev 9 ...
Страница 431: ...GE ...