TuffSat User’s Guide and Service Manual
2-4
Checking normal operation
Before monitoring a patient, always check that the oximeter is operating normally.
WARNING: Patient safety. Never test or perform maintenance on the oximeter while
using it to monitor a patient.
WARNING: Failure of operation.
•
Do not use the oximeter if it fails to function as described or if the validity of
data is questionable. Refer to the appropriate sections of this manual to
identify and correct the malfunction.
•
It is possible for any device to malfunction; therefore, always verify unusual
data by performing a formal patient assessment.
1.
Verify that the four AA batteries are installed. For installation instructions, refer
to chapter 3.
2.
Press the on/off button to power on the oximeter.
The software version number is displayed briefly in the pulse rate display area.
Software version 7.0 is displayed as 070, for example.
As the oximeter completes its self-test, verify the following:
•
The backlight illuminates.
•
All display elements (including the printer and battery screen icons)
illuminate briefly, then all turn off.
•
The backlight remains on until dashes (–––) are displayed for the SpO
2
and
pulse rate values.
NOTE: If the low battery icon flashes on and off continuously, change the
batteries as instructed in chapter 3.
3.
Place a sensor on your finger and connect the sensor cable to the oximeter.
4.
When the SpO
2
and pulse rate values are displayed, verify that the lowest pleth
bar segment remains on while one or more of the other segments flash on and
off.
5.
Press the PI
r
button and verify that the PI
r
pulsatile value is displayed. Press the
PI
r
button again to display the SpO
2
and pulse rate.
6.
Disconnect the sensor cable from the oximeter. Verify that dashes are displayed
in place of the SpO
2
and pulse rate values.
7.
Reconnect the sensor cable to the oximeter. After SpO
2
and pulse rate values are
displayed, remove the sensor from your finger and verify that dashes are
displayed.
If the oximeter fails to function as described, refer to
Troubleshooting
in chapter 3.