OPERATION & MAINTENANCE SPECIFICATIONS MANUAL

09

07

Section:

Subject:

EQUIPMENT/TOOLS

ULTRASONIC THICKNESS MEASUREMENT

Procedure Issue Date: February 2007

Revision

2

Approved by:

Page 9 of 10

\\Stresstel T-Mike Update

Turn on instrument Now

X

CLOSE

5918.2

English



Backlight

Measurement Units

Serial Port

Inches
Millimeters

COM1
COM2

T-Mike Settings

\\Stresstel SetMike

Transmit Settings

Velocity (M/S)

Language

CLOSE

SEND

X

the LCD will indicate the velocity. Now turn on the T-Mike B. The T-Mike B should display CAL, followed
by the new velocity and then will go into the measurement mode. Be sure this new velocity matches with
the sound velocity in the T-Mike E.

5.

To Program from an IBM Compatible PC

Remove rubber plug from the serial port of the T-Mike and insert the LEMO connector of the T-Mike serial
cable (Part Number 022-506-783). Connect the serial cable to COM1 or COM2 of the PC. Run SETMIKE
from Start/Programs/Stresstel or from the appropriate installation directory. Select Measurement Unit:
Inches or Millimeters. Enter sound velocity in inches/uS or M/S. Press

Enter

to accept the new velocity.

Check Backlight to illuminate T-Mike reading. Select Serial Port: COM1 or COM2.

Use the Tab key to cycle through the
SETMIKE Windows application controls

.

Press

SEND

to initiate update of the T-Mike. Turn on the T-Mike. If the T-Mike was already on, turn it off

and turn on gain.

The T-Mike P/B may require several attempts to synchronize
with SETMIKE. Repeat Off/ON cycles until the correct velocity
appears in the display.

When you turn off the T-Mike B, it will display the EPROM software version. When you turn on the T-
Mike B, it will display the active material. Press MAT again to display the sound velocity. Test the T-
Mike B on a material sample to verify that it reads the correct thickness.

E.

HELPFUL HINTS on OPERATION

1.

Backlight

The T-Mike E features a backlight LCD display. The display illuminates when tan echo is detected.
Using the backlight will significantly reduce battery life. Therefore, you may wish to deactivate the
backlight when using the T-Mike E in a well-lit environment.

2.

Clean the Surface

Clean the probe and the material to be material to be measured and apply only enough couplant to
obtain a steady, valid reading. Usually one drop is enough.

Содержание StressTel T-MIKE E

Страница 1: ... exhaustive in the treatment of ultrasonic technique as it is provided for general guidance only SAFETY Prior to excavation for inspection verification shall be made that the pipeline has been established at a safe working pressure in accordance with the requirements of O MSM Section 12 Reference 17 Safe Working Pressure Defect excavation shall then be completed in accordance with the requirements...

Страница 2: ...ended that the backlight be OFF when measurements are made in well illuminated areas as this will greatly increase battery life 3 Metric English Units Press the CAL key a second time The T Mike E will display the units in which it is set to measure thickness IN or MM Press the UP or DOWN arrow keys to select the desired unit of measurement 4 Calibration Lock Press the CAL key a third time The T Mi...

Страница 3: ...Zero Function The PROBE ZERO function compensates the T Mike E for the fixed delay of the ultrasonic transducer This function should be performed at the start of each day to compensate for transducer wear or whenever the transducer is changed Apply a drop of couplant to the end of the transducer Press the ON OFF button on they keyboard to turn on the T Mike E Place the transducer in steady contact...

Страница 4: ...ation mode and enables the UP DOWN arrow keys Adjust the thickness shown on the T Mike E display by using the UP and DOWN arrows item 10 in Figure 2 Note that when these keys are held continuously the speed at which the thickness changes will increase making it easy to make the large adjustments for materials such as plastic and aluminum When the correct thickness appears on the display press the ...

Страница 5: ...y press the CAL key again to enter the velocity for use in thickness measurements and to return to the measurement mode 5 Thickness Measurement Remove dirt loose material and couplant residue from the surface of material at points where measurements are desired Verify that the PROBE and CAL functions have been performed Place a drop of couplant on the material surface at the measurement point Plac...

Страница 6: ...nt charge is available to operate the unit properly When the T Mike E detects minimum allowable battery level the numbers on the display will blink on and off This alerts the operator that the batteries should be replaced or recharged before further operation When the battery level fails below the minimum allowable level the T Mike E will turn off and not allow further operations until batteries a...

Страница 7: ...ocity Turn off both gauges Connect the special cable between the two data charger ports Turn on the T Mike E press the Cal key twice and the LCD will indicate the velocity Now turn on the T Mike Programmable The T Mike Programmable should display CAL followed by the new velocity and then will go into the measurement mode Be sure this new velocity matches with the sound velocity in the T Mike E 3 T...

Страница 8: ...Mike E In either case a special cable is required 2 Material Selection First follow the above PROBE ZERO procedure to calibrate the delay of the ultrasonic transducer being used Press the MAT key to select the material being tested Select the material by using the UP and DOWN arrows Press the MAT key a second time to store the velocity for the selected material Some materials may exhibit sound vel...

Страница 9: ...Enter sound velocity in inches uS or M S Press Enter to accept the new velocity Check Backlight to illuminate T Mike reading Select Serial Port COM1 or COM2 Use the Tab key to cycle through the SETMIKE Windows application controls Press SEND to initiate update of the T Mike Turn on the T Mike If the T Mike was already on turn it off and turn on gain The T Mike P B may require several attempts to s...

Страница 10: ...ece are the same temperature shape material and size NOTICE Since velocity of sound varies from test piece to work piece accuracy of measurement depends on consistency of ultrasonic velocity Ultrasonic velocity also varies with internal stress so that heat treatment of the material will also affect accuracy The following table expresses accuracy variations for some common materials NOTE These spec...

Отзывы: