![GE StressTel T-MIKE E Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/ge/stresstel-t-mike-e/stresstel-t-mike-e_operation-and-maintenance-specifications-manual_128450010.webp)
Section No.
Reference No.
OPERATION & MAINTENANCE SPECIFICATIONS MANUAL
09
07
Section:
Subject:
EQUIPMENT/TOOLS
ULTRASONIC THICKNESS MEASUREMENT
Procedure Issue Date: February 2007
Revision
2
Approved by:
Page 10 of 10
3.
Material Consistency
Maximum measurement accuracy will be contained if the work piece and the calibration piece are the
same temperature, shape, material, and size.
NOTICE
Since velocity of sound varies from test piece to work piece, accuracy of measurement depends on
consistency of ultrasonic velocity. Ultrasonic velocity also varies with internal stress, so that heat
treatment of the material will also affect accuracy. The following table expresses accuracy variations for
some common materials. NOTE: These specifications are intended only as a general guide. Actual
accuracy may be different.
Aluminum
+/- 2%
Cast iron
+/- 8%
Steel
+/- 0.5%
Nylon
+/- 10%