CABLE CONCEALMENT
Front of magnet
NOTE:
This drawing is to be used only as a design intent document. refer to ge installation manual for tray install.
actual tray installation may be site dependent.
This drawing not to scale
Customer/contractor installed
cable concealment frame
customer/contractor to provide
opening and install frame.
Magnet
geometric
isocenter
Cable trays
200
[7.9 in
]
700
[27.6 in
]
673
[26.5 in]
1355
[53.3 in]
673
[26.5 in]
Vibromat
NOT TO SCALE
MAGNET ON VIBROACOUSTIC DAMPENING KIT "VIBROMAT"
Vibromat
Patient table
dock frame
anchor hole
seismic anchor holes
4 x magnet
mounting holes
VibroAcoustic Mat (M7000VA)
weight: 8 kg (each) [17 lbs]
390
[15.4 in]
MAGNET ROOM FLOOR SPECIFICATIONS
The finished floor must support the weight of all components (e.g., patient table, gradient coil replacement cart)
throughout operation and service life.
Floor must be flat and level to 3 mm [0.12in] between high and low spots over the area shown.
3419mm
[134.61in]
943mm
[37.13in]
PATIENT TABLE DOCK ANCHOR MOUNTING REQUIREMENTS
The RF Shield vendor must design and install the dock anchor bolt
The dock anchor hole must be drilled after the Magnet is installed
The dock anchor must not contact floor rebar or other structural steel
The dock anchor must electrically contact the RF shield at point of entry
The RF shield vendor must perform a pull test on the anchor (equal to the clamping force).
NOT TO SCALE
FEMALE ANCHOR METHOD
(CUTAWAY SIDE VIEW)
BOLTED METHOD
(CUTAWAY SIDE VIEW)
1
Removable Anchor Rod (Male insert)
2
Dock
3
Clamp bracket
4
Finished floor
5
Filler Board or Grout
6
RF Shield
7
Conductive Fibrous Washer (RF seal)
8
Concrete
9
Female Anchor Insert
1
2
3
4
5
6
7
8
9
8
7
6
5
4
3
2
1
Date
Rev
/24
EN-MRI-TYP-SIGNA_PIONEER-WEB.DWG
A
23/Jul/2018
SIGNA PIONEER
Typical
S3 - Structural Details
13