Chapter 6: Analyzer Operation
Continuous Measurements
By turning
Continuous
On
, the Analyzer makes continuous measurements of the enabled sample streams and
the Check Standard (if it is enabled). Continuous measurement of Stream 1 is shown in
. Only Stream 1
and the Check Standard are enabled in this example, so measurements of Stream 1 are made without a pause,
with two exceptions.
Figure 41: Continuous Operation
The first exception is the Check Standard, since it is enabled. Measurements of the Check Standard are made
according to the schedule established by the operator. In the example shown in
, the check standard is
measured every three hours (the Check Standard Frequency). Since two
Repetitions
have been selected, the
check standard is measured twice every three hours. The first measurement is ignored because it has been
designated as a
Reject
.
Figure 42: Operation with a Cycle Frequency of One Hour
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...