K177 Issue No. 3
10
RPT 301 User Manual
connect screen
to connector shell
screened cable
computer
enclosure
RPT
301
Screening to ensure EMC performance
Screening
Make sure that the chassis connection between the control computer
and the instrument is continuous. Use screened cable and bond the
cable screening to the connector shell of the instrument and to the
computer enclosure.