MAINTENANCE
GE Medical Systems
SENOGRAPHE 800T
REV 6
om 2135575–100
3–28
SECTION 5
CENTERING DEVICE LIGHT BULB REPLACEMENT
When the light bulb of the centering device fails it can be easily replaced by the user following
the instructions below. This operation takes only one minute (after the bulb has cooled
down).
Use the spare bulb in the first–aid kit.
You are advised to order a new replacement bulb now, from your local GE After Sales Service
or by using the accessory catalogues (DIA, Numéris).
THE BULB AND ITS PROTECTION COVER MAY BE VERY HOT. LEAVE
THE CENTERING DEVICE SWITCHED OFF TO COOL DOWN FOR
TWENTY (20) MINUTES BEFORE REMOVING THE BULB.
1. Switch off the Senographe 800T.
2. At the back of the collimator, remove the cover (1) by pulling on the handle at the bottom
(2).
X
2
1
WARNING
Содержание SENOGRAPHE 800T
Страница 6: ...GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 iv Blank page ...
Страница 12: ...GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 x Blank page ...
Страница 14: ...GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 xii Blank page ...
Страница 16: ...GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 xiv Blank page ...
Страница 33: ...INTRODUCTION GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 1 17 Blank page ...
Страница 38: ...INTRODUCTION GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 1 22 Blank page ...
Страница 65: ...MAINTENANCE GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 3 21 Confirm functions selection ...
Страница 80: ...SPECIFICATIONS GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 4 6 Blank page ...
Страница 82: ...REGULATIONS GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 5 2 Blank page ...