MAINTENANCE
GE Medical Systems
SENOGRAPHE 800T
REV 6
om 2135575–100
3–25
4–7
Exposure Interlock Disable
The Senographe 800T is equipped with an exposure interlock feature which prevents an
exposure from being made if there is no cassette in the Bucky, or if the cassette from the
previous exposure has not been changed.
Trying to make an exposure without a cassette in the Bucky or cassette holder, or without
having changed the cassette between two exposures will result in a beep sound and the red
indicator light located above the Bucky will come ON. The exposure will be inhibited.
However, it is possible, for special purposes such as physicist testing, to disable this interlock.
Note:
The Senographe 800T AUTOMATICALLY enables exposure interlock when powered
on.
4–7–1
Access to Programming Mode
4–7–2
Selecting a Function
To select a function, press the key located under the symbol for that function.
Содержание SENOGRAPHE 800T
Страница 6: ...GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 iv Blank page ...
Страница 12: ...GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 x Blank page ...
Страница 14: ...GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 xii Blank page ...
Страница 16: ...GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 xiv Blank page ...
Страница 33: ...INTRODUCTION GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 1 17 Blank page ...
Страница 38: ...INTRODUCTION GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 1 22 Blank page ...
Страница 65: ...MAINTENANCE GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 3 21 Confirm functions selection ...
Страница 80: ...SPECIFICATIONS GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 4 6 Blank page ...
Страница 82: ...REGULATIONS GE Medical Systems SENOGRAPHE 800T REV 6 om 2135575 100 5 2 Blank page ...