4.3.1.2 kV Assist kV Mode
When
kV Assist
Mode is selected, the system will adjust kV and NI / minimum and maximum
mA based on the patient attenuation from the scout, and the protocol’s clinical task, baseline kV,
kV range, mA mode, and mA setting or mA range and Noise Index.
See the
Protocol
chapter for additional details on setting up kV Assist in protocols.
4.3.2 mA Mode
From the
kV and mA Control
collection, select
mA Mode
.
The mA value determines the focal spot size that can be used to limit the system to a small
focal spot. You need to enter the correct mA value for the focal spot you wish to use based on
and
. The system minimum is 10 mA. The system maximum for SmartPrep
scans is 140 mA. The system maximum for scout scans is 250 mA. The system maximum for
pediatric scans is kV dependent based on a limit of 42 kW. For Helical scans, the mA may be
further constrained by the duration of the scan.
Table 17: Maximum permissible mA values for single 0.3 second exposure
Single Scan mA Ratings: 0.3s
Focal Spot
Scan Mode
kV
Extra Large
Large
Small
Normal
70
375
375
250
80
580
580
580
100
720
720
490
120
740
665
405
140
635
570
350
Hi-Res
70
N/A
375
250
80
N/A
580
425
100
N/A
545
340
120
N/A
455
280
140
N/A
390
240
Table 18: Maximum permissible mA values for single 4 second exposure
Single Scan mA Ratings: 4s
Focal Spot
Scan Mode
kV
Extra Large
Large
Small
Normal
70
375
375
250
80
580
580
550
100
720
660
440
120
605
550
365
140
520
470
310
Hi-Res
70
N/A
375
250
80
N/A
580
385
100
N/A
475
310
120
N/A
395
255
Revolution CT User Manual
Direction 5480385-1EN, Revision 1
Chapter 11 Scan
271
Содержание Revolution CT
Страница 1: ...Revolution CT User Manual OPERATING DOCUMENTATION 5480385 1EN Revision 1 ...
Страница 2: ...This page left intentionally blank Revolution CT User Manual Direction 5480385 1EN Revision 1 2 ...
Страница 216: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 216 2 Patient Schedule ...
Страница 370: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 370 7 SnapShot Freeze ...
Страница 440: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 440 4 Exam Split ...
Страница 471: ...Illustration 18 VR Settings screen Revolution CT User Manual Direction 5480385 1EN Revision 1 Chapter 16 Reformat 471 ...
Страница 502: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 502 7 Batch Film ...
Страница 674: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 674 1 Operator Messages ...
Страница 680: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 680 1 Units of Measure ...
Страница 688: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 688 1 Lexicon ...
Страница 692: ...Illustration 2 GE XXL Revolution CT User Manual Direction 5480385 1EN Revision 1 692 1 Auto Gating Profiles ...
Страница 694: ...Illustration 4 GE Gen CardioVasc Revolution CT User Manual Direction 5480385 1EN Revision 1 694 1 Auto Gating Profiles ...
Страница 696: ...Illustration 6 GE CACS Revolution CT User Manual Direction 5480385 1EN Revision 1 696 1 Auto Gating Profiles ...
Страница 701: ...Revolution CT User Manual Direction 5480385 1EN Revision 1 This page left intentionally blank 701 ...