9-9
ILLUSTRATION 9-6
Bucky Angulation
1.
Press one of pushbuttons located behind the bucky support
2.Keeping the push-button pressed, angulate the bucky to the desired position.
3.Release the push-button in order the bucky to remain locked.
SECTION 5
PERIODIC MAINTENANCE
5-1 Maintenance Information
In order to assure continued safe performance of x-ray equipment, a periodic
maintenance program must be established. It is the owner's responsibility to sup-
ply or arrange for this. After installation, maintenance should be performed on the
SG80/120 every eighteen months. The equipment should be cleaned frequently,
particularly if corroding chemicals are present. Clean stainless steel surfaces
(patient support side rails) with a plastic scouring pad (Scotchbrite or similar).
Clean other surfaces with a cloth moistened in warm water with mild soap. Wipe
with a cloth moistened in clean water.
DO NOT USE CLEANERS OR SOLVENTS OF ANY KIND AS THY MAY
DULL THE FINISH OR BLUR THE LETTERING. POLISH WITH CLEAN
WATER OR PASTE WAX. DO NOT US A WAX CONTAINING A
CLEANING SUBSTANCE.
NEVER ATTEMPT TO CLEAN ANY PART OF THE TABLE WHEN THE
SG80/120 IS POWERED “ON”. SWITCH OFF LINE POWER SUPPLY
AND THE X-RAY GENERATOR BEFORE CLEANING.