K0450 Revision B
2
[EN] English
5.1 PACE Controller Connection
WARNING
ON COMPLETION OF CALIBRATION, OPEN THE ON/OFF VALVE TO VENT TRAPPED
PRESSURE FROM THE PORT TO ATMOSPHERE.
Connect the calibrator for each pressure range as follows:
Connect the output of the pressure standard to the PACE module output port.
Fit an on/off valve to the input of the PACE controller. Leave the other side of the
valve open to atmosphere. Turn the valve to the closed position.
Gauge Reference
If the pressure standard has a reference connection, then connect this to the PACE reference
port on the module manifold. Otherwise the calibrator reference port should be open to
atmosphere.
Note:
Module 2 reference port open to atmosphere or to reference port of the pressure
standard.
Note:
Optimum performance is achieved by connecting the PACE reference port to the
pressure standard.
SUPPLY
OUTLET VENT
REF
+
-
Module 2
SUPPLY
OUTLET VENT
REF
+
-
Module 1
PACE Controller
Rear view of
instrument
Pressure Standard
On/off valve fitted to