E4H Series
Operation & Maintenance Manual
43
2. Press 1 or highlight OPERATION and press ENT. There may be a dark flashing line separating
the two sections; this indicates which alarms are active at the moment. As shown at the
bottom of the screen, press ENT to access the relays.
3. The four relays line up vertically with boxes that are blank when the relay is not in operation.
Select a relay by pressing its number. The box will change (it will probably become shaded),
indicating that the relay has reversed its status.
4. NOTE that relays retain manual mode even after powering down.
5. Finally, press CLR twice to return to the MAIN MENU.
A.1.6 Calibration
Methodology
1. You must have an accurate reading of the system water to properly calibrate the monitor.
•
A properly calibrated handheld temperature, conductivity, or pH meter may be acceptable for
completing calibrations
•
NOTE:
If at any time you believe that the pH, conductivity, or temperature calibrations have
been improperly performed, you may reinitialize the calibrations and restore them to their
default values.
2. Calibrate the temperature first.
•
Measure the temperature of the system water
•
Press PRO on the keypad, select ZERO
•
Use the number keys to enter the value, then press ENT
•
Do not choose SPAN when calibrating temperature
3. To calibrate pH
•
Measure the pH of the system water with a handheld meter
•
Press PRO on the keypad, select zero
•
Use the number keys to enter the value, then press ENT
•
Do not choose SPAN when calibrating pH
4. To calibrate conductivity
•
Take a system water sample reading with your handheld meter
•
Press PRO on the keypad
•
Use the number keys to enter the value, then press ENT
•
When the number is accepted, you will see the SPAN COMPLETE screen before the LCD display
switches back to the original OPERATION screen
5.
Take a second sample and confirm that the readings on the monitor
display are correct.
A.1.7
Setpoints (pH, Conductivity, Temperature, Permeate and Concentrate Flow)
Typically, pH is the only process that will be controlled. Follow these instructions to establish the
monitor’s setpoint.
1. From the MAIN MENU, press 2 or highlight RELAYS and press ENT.
2. Press the number of the relay governing the operation you wish to control.
3. Press 2 or highlight SETPOINT CONTROL and press ENT.
Содержание Osmonics E4H Series
Страница 36: ......
Страница 39: ...E4H Series Operation Maintenance Manual 39 1 2450 Front Panel with Display ...
Страница 45: ...E4H Series Operation Maintenance Manual 45 A 6 Technician Level Menus ...
Страница 46: ...E4H Series Operation Maintenance Manual 46 A 1 8 Operation ...
Страница 47: ...E4H Series Operation Maintenance Manual 47 ...
Страница 48: ...E4H Series Operation Maintenance Manual 48 ...
Страница 49: ...E4H Series Operation Maintenance Manual 49 A 1 9 Relays ...
Страница 50: ...E4H Series Operation Maintenance Manual 50 A 1 10 Alarms ...
Страница 51: ...E4H Series Operation Maintenance Manual 51 A 1 11 Flow Meters ...
Страница 52: ...E4H Series Operation Maintenance Manual 52 A 1 12 4 20 mA Output ...
Страница 53: ...E4H Series Operation Maintenance Manual 53 A 1 13 System Setup ...
Страница 54: ...E4H Series Operation Maintenance Manual 54 ...
Страница 55: ...E4H Series Operation Maintenance Manual 55 ...
Страница 56: ...E4H Series Operation Maintenance Manual 56 ...
Страница 57: ...E4H Series Operation Maintenance Manual 57 A 1 14 Clock ...
Страница 58: ...E4H Series Operation Maintenance Manual 58 A 7 Security Level Menus ...
Страница 59: ...E4H Series Operation Maintenance Manual 59 A 8 Operator View Level Menus ...