Section 3.0 - Console Connections
-
Select keyboard overlay from keyboard collector (5324605-XX) to attach. (See
Figure 2-4 Keyboard with the English film and Warning Label Installed
-
Select the GSCB overlay (with both Tilt and E-Reset, P/N is 5401237-XXX) install the
proper overlay with the appropriate language for the system.
Verify that none of the buttons get caught and stuck under the overlay. Pay close attention to the
prescribed tilt button on systems with the tilt feature.
Note:
X-ray ON sound can be turned off / on using the switch on GSCB bottom if customer does not like
it and if local regulation does not require X-ray ON sound. Detail information refer to
Service
Methods->Troubleshooting->Console-> GSCB Troubleshooting
.
4.) The keyboard should attach to the GSCB using the supplied Velcro strip and fit snugly against
the GSCB when finished, as shown in
.
5.) Route the GSCB cable and connect connectors according to
and
Filming
Window / Level
Film
Image
Film
Screen
Film
MID
Series
Previous
W/L
Abdomen Head
Lung
Media-
stinum
Spine
Vertebra
AutoFilm
Image