Configuration
2026419-033E
CIC Pro™
6-7
NOTE
In the hierarchy of multiple compatible CIC Pro center hardware and
software versions that co-exist, it is important to assign a set of the highest
MC IP addresses to the highest software version. For example, if you are
installing CIC v5.1.x to an existing CARESCAPE Network comprising of
CIC v4.0.7 or later, you must allocate a set of the highest MC IP addresses to
all of the CIC Pro center’s hardware running CIC v5.1.x. No lower version
CIC Pro center can have an IP address on the MC network that is greater
than the IP address of a higher version CIC Pro center.
10. Verify that there is
only one
Time Master on the network and it is compliant with
step
by running the following command from any CIC Pro center currently on
the network:
lw -s “TIME MASTER”(lw<space>-s<space>”TIME<space>MASTER”)
11. If you have not already connected the service PC to the CIC Pro center, set up the
service PC’s network properties on page
12. Proceed to the
Configuration process checklist on page 6-7
Configuration process checklist
NOTE
After making configuration changes, ensure the CIC Pro center is functioning
properly by completing all relevant checkout procedures
before
returning it to
See Checkout procedures on page 7-1.
NOTE
If the System Resource Indicator (
icon in the bottom tray) turns yellow or
red during or after configuration, a pop-up message will appear. Take the
necessary action. Refer to
Environment Monitor messages on page 8-19
.
Disconnect CARESCAPE Network IX and MC networks
Disconnect from the CARESCAPE Network IX and MC networks on page 6-10
Install licenses
Activate software licenses on page 6-10
.
Configure Webmin-related settings
NOTE
All Webmin modules are static in nature. Since the Webmin service interface
session times out after 15 minutes, you must always refresh the browser to load
the latest page.
Содержание MP100 Series
Страница 12: ...x CIC Pro 2026419 033E...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Страница 313: ......