REPLACEMENT OF THE PREFILTER, POSTFILTER, AND RO MEMBRANE ELEMENTS
Rev D
6. Remove exhausted prefilter and discard.
7. If Changing Membrane Elements:
A. Remove membrane sumps. Remove and discard used
elements.
B. Remove new elements from packaging.
C. Lubricate element O-rings, brine seals, and sump O-rings
with silicone lubricant.
D. Securely insert O-ring end of elements into manifold. See
Figure 8.
E. Replace sumps and tighten until it bottoms out.
8. Sanitize the system.
A. The manifold should be positioned flat with the sump
connections facing up.
Figure 8
CAUTION:
The person handling the filters and membrane elements
must have clean hands to keep the system sanitized. The use of
sterile/latex gloves is recommended.
CAUTION:
Elements contain a foodgrade preservative. The use of
sterile/latex gloves is highly recommended.
NOTE:
The system should be sanitized whenever a membrane
element or filter is replaced.
Prefilter Center Opening
Membrane Element Locations