Electrical Diagram for Patient Leakage Current Test
For the protection of the test person, the following values of resistor R may be used:
Type BF
22 kOhm (120 to 130 V)
47 kOhm (220 to 240 V)
Type CF
100 kOhm (220 to 240 V)
7.2 Maintenance, cleaning, disinfection
MAC 1200 maintenance, cleaning, disinfection is performed in accordance with the
MAC 1200 “Operator’s Manual”, Section 11, "Cleaning, disinfection and maintenance",
as applicable.
Applications requiring extensive recordings may result in deposits on the thermal array
printhead which normally do not, however, have any adverse effect on the printing
quality. This can be removed with a soft, fluff-free cloth soaked in an alcohol-based
cleaning agent (e.g., surgical spirit).
Содержание marquette MAC 1100
Страница 1: ...MAC 1100 1200 Servicing Instructions Version 1 1 227 492 20 SA e Revision D...
Страница 6: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 6 227 492 20 D 0002...
Страница 40: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 40 227 492 20 D 0002...
Страница 44: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 44 227 492 20 D 0002...
Страница 48: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 48 227 492 20 D 0002...
Страница 50: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 50 227 492 20 D 0002...
Страница 62: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 62 227 492 20 D 0002...
Страница 70: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 70 227 492 20 D 0002...
Страница 72: ...Marquette Hellige GmbH MAC 1100 1200 V 1 1 Page 72 227 492 20 D 0002...
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 86: ......
Страница 87: ......
Страница 89: ......
Страница 91: ......