Revision L
MAC™ 3500 Resting ECG Analysis System
4-27
2021337-036
Troubleshooting:
Equipment Problems
Equipment Problems
ECG Data Noise
If the acquired ECG data displays unacceptable noise levels:
When troubleshooting noise or signal quality, be sure the problem is not being
caused by poor skin preparation, or placement and condition of electrodes.
Careful skin preparation is the key to an interference-free ECG. Signal quality is
indicated using Hookup advisor. Click here to learn more about Hook-up
Advisor. Hookup Advisor can be turned on or off in the ECG menu.
Select
System Setup
ECG
ECG Acquisition
.
Check for defective or date expired electrodes.
Check for defective, broken, or disconnected leadwires.
Run the Acquisition Module Tests in the Diagnostic menu and make sure all
lead wires pass the noise test.
Содержание MAC 3500
Страница 1: ...GE Healthcare MACTM 3500 Resting ECG Analysis System Version 9D Service Manual 2021337 036 Revision L ...
Страница 9: ...Revision L MAC 3500 Resting ECG Analysis System 1 1 2021337 036 1 Introduction ...
Страница 10: ...1 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 19: ...Revision L MAC 3500 Resting ECG Analysis System 2 1 2021337 036 2 Equipment Overview ...
Страница 20: ...2 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 25: ...Revision L MAC 3500 Resting ECG Analysis System 3 1 2021337 036 3 Installation ...
Страница 26: ...3 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 40: ...3 16 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Installation General Assembly For your notes ...
Страница 41: ...Revision L MAC 3500 Resting ECG Analysis System 4 1 2021337 036 4 Troubleshooting ...
Страница 42: ...4 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 44: ...4 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Troubleshooting Assembly Descriptions ...
Страница 83: ...Revision L MAC 3500 Resting ECG Analysis System 5 1 2021337 036 5 Maintenance ...
Страница 84: ...5 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 122: ...5 40 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Maintenance Functional Checkout Procedures ...
Страница 123: ...Revision L MAC 3500 Resting ECG Analysis System 6 1 2021337 036 6 Parts Lists ...
Страница 124: ...6 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 152: ...6 30 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Parts Lists Field Replaceable Units FRUs ...
Страница 153: ...Revision L MAC 3500 Resting ECG Analysis System A 1 2021337 036 A Technical Specifications ...
Страница 154: ...A 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 160: ...A 8 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Technical Specifications Barcode Scanner For your notes ...
Страница 161: ...Revision L MAC 3500 Resting ECG Analysis System B 1 2021337 036 B Software Hardware Compatibility ...
Страница 162: ...B 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 167: ...Revision L MAC 3500 Resting ECG Analysis System C 1 2021337 036 C Electromagnetic Compatibility ...
Страница 168: ...C 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 175: ...Revision L MAC 3500 Resting ECG Analysis System Index 1 2021337 036 Index ...
Страница 176: ...Index 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 178: ...Index Index 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 ...
Страница 179: ......