Maintenance
5.
If it was installed, replace the KISS pump and tubing.
See step
through step
in
“Replacing the KISS Pump Assembly” on page 127
for instructions.
6.
Replace the power supply assembly.
See step
through step
in
“Replacing the Power Supply Assembly” on page
for instructions.
7.
Reattach the top cover assembly.
See
“Reattaching the Top Cover Assembly to the Device” on page 81
for
instructions.
8.
Reinsert the battery.
See step
through step
in
“Replacing the Battery” on page 74
for instructions.
9.
Continue with the appropriate functional checkout procedure for this FRU.
See
“Functional Checkout” on page 129
for more information.
Removing the Chassis
Use the following procedure to remove the chassis from the device in order to replace
it. For instructions on how to pull the chassis in order to access the printer or to
remove the bottom cover, see
“Detaching the Chassis” on page 82
1.
Remove the top cover assembly, if it is has not been removed already.
See
“Removing the Top Cover Assembly” on page 75
for more information.
2.
Remove the power supply assembly.
See step
through step
in
“Removing the Power Supply Assembly” on page
for instructions.
3.
If it is installed, remove the KISS pump and tubing.
See step
through step
in
“Removing the KISS Pump Assembly” on page 126
for instructions.
4.
Remove the fasteners securing the power supply/KISS pump base to the chassis,
as identified in the following photograph.
86
MAC™ 2000 ECG Analysis System
2053535-003C
Содержание MAC 2000
Страница 41: ...Product Overview 2053535 003C MAC 2000 ECG Analysis System 41 ...
Страница 134: ...Maintenance 134 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 160: ...Parts List 160 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 182: ...182 MAC 2000 ECG Analysis System 2053535 003C ...
Страница 183: ......